Block mounting devices for linacs--drawbacks and modifications: Shielding blocks are often used in radiotherapy. They are fixed on the linac's head by means of standard accessory mounting devices. The disadvantages of these parts are discussed and two modifications are presented: One device with adjustable focus-support distance for standard lead blocks and another with gliding planes to insert easily the heavy individually formed shielding blocks. Errors in mounting of all these devices are very dangerous, so we developed a simple locking lever providing a high security.
直线加速器的挡块安装装置——缺点与改进:屏蔽挡块常用于放射治疗。它们通过标准附件安装装置固定在直线加速器的机头。本文讨论了这些部件的缺点,并提出了两种改进方案:一种是用于标准铅挡块的焦点支撑距离可调装置,另一种是带有滑动平面以便轻松插入重型定制屏蔽挡块的装置。所有这些装置的安装误差都非常危险,因此我们开发了一种提供高安全性的简单锁定杆。