骶前间隙及其对骶神经调节器植入的影响。

The presacral space and its impact on sacral neuromodulator implantation.

作者信息

Saint Clair Nicole, Boyles Sarah Hamilton, Clark Amanda, Edwards S Renee, Denman Mary Anna, Gregory W Thomas

机构信息

Oregon Health & Science University, Portland, Oregon 97239, USA.

出版信息

J Urol. 2008 Sep;180(3):988-91. doi: 10.1016/j.juro.2008.05.032. Epub 2008 Jul 17.

Abstract

PURPOSE

We describe the presacral space and its potential impact on sacral neuromodulator implantation and bowel injury.

MATERIALS AND METHODS

Parasagittal images containing bilateral sacral foramina (S2-S4) were examined on 45 pelvic magnetic resonance images. Images were excluded from analysis if they were poor quality or had any history causing distortion of normal anatomy. We measured the natural angle between the foramina and the dorsal skin to approximate the needle angulation during neuromodulator electrode placement. Using these angles we measured the distance from the skin to any bowel (D1), the skin to the dorsal sacrum (D2) and then calculated the distance from the dorsal sacrum to any bowel (D3).

RESULTS

Mean subject age was 45 years (range 19 to 78) and body mass index was 27.9 kg/m(2) (range 18.6 to 56.2). At S3 the mean foraminal angle and D3 were 46 +/- 8.4 degrees and 27.4 +/- 11.7 mm, respectively. Increasing age was moderately correlated to widening D3 at each foramina (r = 0.3, Pearson's p <0.05). Body mass index did not consistently vary with D3 at any foramina.

CONCLUSIONS

Our measurements suggest that the presacral space can be expected to be approximately 27 mm at the level of S3 where the neuromodulator electrode is implanted. It is possible to encounter bowel while performing this implantation using standard techniques and equipment. We recommend the standard use of fluoroscopy during placement.

摘要

目的

我们描述了骶前间隙及其对骶神经调节器植入和肠道损伤的潜在影响。

材料与方法

在45例盆腔磁共振图像上检查包含双侧骶孔(S2 - S4)的矢状旁位图像。如果图像质量差或有任何导致正常解剖结构变形的病史,则将其排除在分析之外。我们测量了骶孔与背部皮肤之间的自然角度,以近似神经调节器电极放置过程中的进针角度。利用这些角度,我们测量了皮肤到任何肠道的距离(D1)、皮肤到骶骨背侧的距离(D2),然后计算了骶骨背侧到任何肠道的距离(D3)。

结果

受试者平均年龄为45岁(范围19至78岁),体重指数为27.9 kg/m²(范围18.6至56.2)。在S3水平,平均骶孔角度和D3分别为46±8.4度和27.4±11.7毫米。年龄增长与每个骶孔处D3增宽呈中度相关(r = 0.3,Pearson检验p <0.05)。在任何骶孔处,体重指数与D3均无一致变化。

结论

我们的测量结果表明,在植入神经调节器电极的S3水平,骶前间隙预计约为27毫米。使用标准技术和设备进行此植入操作时可能会碰到肠道。我们建议在放置过程中常规使用荧光透视。