Laflamme G Y, Alami G B, Zhim F
Programme d'Orthopedie Edouard Samson (POES), University of Montreal, Montreal, Quebec, Canada.
Hip Int. 2008 Jan-Mar;18(1):29-34. doi: 10.1177/112070000801800106.
In acetabular revisions, polyethylene (PE) liners are often cemented into metal shells on top of acetabular screw heads. This study investigates the possibility of using this technique to obtain fixed-angle acetabular screws Eth a concept that has not yet been reported in the literature. Two groups of screws (n=8) were inserted into Trabecular Metal revision shells (Zimmer), into which PE liners were then cemented. Screws in Group 1, inserted in the shell's pre-fabricated holes, were countersunk, whereas screws in Group 2 were inserted in custom-drilled holes that make their heads protrude into, and interdigitate with, the overlying cement mantle. Perpendicular loading was then applied to the screw shafts both statically to failure and cyclically. A greater stiffness was observed for the protruding screws upon static loading; and while the countersunk screws all failed at the screw-cement junction (53.44 + or - 8.33 N), the protruding screws all failed at the screw shaft (1049.79 + or - 32.12 N) a 20-fold difference (p< 0.05). Under cyclic loading, only the protruding screw head specimen did not fail, undergoing an overall displacement within the limits of osseointegration.These results support the hypothesis that the protrusion of an acetabular screw head into an overlying cement mantle significantly increases its angular stability. Provided other variables are favorable, this locking effect may increase the initial stability of the whole implant, thus improving the ultimate success of complex acetabular revisions.
在髋臼翻修手术中,聚乙烯(PE)内衬通常用骨水泥固定在髋臼螺钉头部上方的金属髋臼杯内。本研究探讨了使用该技术获得固定角度髋臼螺钉的可能性,这一概念在文献中尚未有报道。将两组螺钉(n = 8)插入小梁金属髋臼翻修杯(捷迈公司)中,然后在其中用骨水泥固定PE内衬。第1组螺钉插入髋臼杯预制孔中并埋头,而第2组螺钉插入定制钻孔中,使其头部突入并与覆盖的骨水泥套相互交错。然后对螺钉轴施加垂直载荷,包括静态加载直至破坏和循环加载。静态加载时,突出螺钉表现出更大的刚度;埋头螺钉均在螺钉 - 骨水泥结合处失效(53.44 ± 8.33 N),而突出螺钉均在螺钉轴处失效(1049.79 ± 32.12 N),相差20倍(p < 0.05)。在循环加载下,只有突出螺钉头部试件未失效,在骨整合范围内发生整体位移。这些结果支持以下假设:髋臼螺钉头部突入覆盖的骨水泥套可显著提高其角度稳定性。如果其他变量有利,这种锁定效应可能会增加整个植入物的初始稳定性,从而提高复杂髋臼翻修手术的最终成功率。