Suppr超能文献

一种用于测量胃内pH值的Bravo无线pH监测胶囊的新型放置方法。

A novel placement method of the Bravo wireless pH monitoring capsule for measuring intragastric pH.

作者信息

Chang Jae Hyuck, Choi Myung Gyu, Yim Dong-Seok, Cho Yu Kyung, Park Jae Myung, Lee In Seok, Kim Sang Woo, Chung In Sik

机构信息

Division of Gastroenterology, Department of Internal Medicine, College of Medicine, Kangnam St. Mary's Hospital, The Catholic University of Korea, No. 505, Banpo-Dong, Seocho-Gu, Seoul, 137-040, South Korea.

出版信息

Dig Dis Sci. 2009 Mar;54(3):578-85. doi: 10.1007/s10620-008-0399-3. Epub 2008 Jul 23.

Abstract

PURPOSE

The delivery system of the Bravo capsule was designed for placement on the esophagus. We evaluated the feasibility of our novel placement method of the Bravo capsule using a clip to monitor intragastric pH and to compare the accuracy of the Bravo wireless system to the traditionally used Slimline catheter-Mark III Digitrapper pH monitoring system.

METHODS

The Bravo capsule was placed by clip or conventional delivery system using suction on the gastric wall in 20 fasted subjects. A separate group of ten healthy volunteers underwent simultaneous intragastric pH monitoring for comparison of the two systems with meals.

RESULTS

Early dislodgment rate of the capsules was lower when placed using clipping (20%) than using conventional delivery system (70%) within 48 h after placement. We observed prominent movement of one catheter in the stomach during the study. Post-test calibration drifts of the catheters at pH 7.01 were significantly greater than those of the Bravo capsules (P = 0.02).

CONCLUSION

Our novel clipping method of the Bravo pH capsule placement provided accurate monitoring of intragastric pH with merits of tolerability, acid stability, and fixing position.

摘要

目的

Bravo胶囊输送系统设计用于放置在食管上。我们评估了使用夹子放置Bravo胶囊这种新方法监测胃内pH值的可行性,并比较了Bravo无线系统与传统使用的Slimline导管 - Mark III Digitrapper pH监测系统的准确性。

方法

在20名禁食受试者中,通过夹子或使用传统输送系统在胃壁上进行抽吸来放置Bravo胶囊。另一组10名健康志愿者同时进行胃内pH监测,以比较这两种系统在进食时的情况。

结果

放置后48小时内,使用夹子放置胶囊的早期移位率(20%)低于使用传统输送系统(70%)。在研究过程中,我们观察到一根导管在胃内有明显移动。pH值为7.01时,导管的测试后校准漂移明显大于Bravo胶囊(P = 0.02)。

结论

我们采用的Bravo pH胶囊放置新夹法能准确监测胃内pH值,具有耐受性、酸稳定性和固定位置的优点。