Suppr超能文献

磁共振成像期间人工耳蜗磁铁移位

Cochlear implant magnet displacement during magnetic resonance imaging.

作者信息

Deneuve Sophie, Loundon Nathalie, Leboulanger Nicolas, Rouillon Isabelle, Garabedian Erea Noel

机构信息

ENT Department-Hôpital d'Enfants Armand Trousseau, Paris, France.

出版信息

Otol Neurotol. 2008 Sep;29(6):789-190. doi: 10.1097/MAO.0b013e3181825695.

Abstract

OBJECTIVE

Magnet displacement is a rare but possible complication in patients with cochlear implants. We report 1 case in a young child that occurred during magnetic resonance imaging scanning, despite precautionary measures taken by the surgeon and the radiographer.

STUDY DESIGN

Retrospective case report.

SETTING

This case is presented by the ENT Department of Armand Trousseau Paediatrics Hospital, Paris, France.

PATIENT

An 8-year-old child, implanted with a Freedom Contour Nucleus cochlear implant, was referred for a 1.5-T cerebral magnetic resonance imaging because of a progressive neurologic disorder.

CONCLUSION

This rare complication underlines the importance of risk information and preventive measures required, even in case of compatible devices, for performing a magnetic resonance imaging examination in patients wearing a cochlear implant with removable magnet.

摘要

目的

磁体移位是人工耳蜗植入患者中一种罕见但可能出现的并发症。我们报告1例发生在幼儿身上的病例,该病例发生在磁共振成像扫描期间,尽管外科医生和放射技师采取了预防措施。

研究设计

回顾性病例报告。

背景

本病例由法国巴黎阿曼德·特鲁索儿科医院耳鼻喉科提供。

患者

一名8岁儿童,植入了自由轮廓核型人工耳蜗,因进行性神经系统疾病接受1.5-T脑磁共振成像检查。

结论

这种罕见的并发症凸显了风险告知和所需预防措施的重要性,即使使用兼容设备,对于佩戴带有可移除磁体的人工耳蜗的患者进行磁共振成像检查时也是如此。