Suppr超能文献

金属对金属髋关节假体的摩擦学与磨损:髋臼杯角度和股骨头位置的影响

Tribology and wear of metal-on-metal hip prostheses: influence of cup angle and head position.

作者信息

Williams Sophie, Leslie Ian, Isaac Graham, Jin Zhongmin, Ingham Eileen, Fisher John

机构信息

Institute of Medical and Biological Engineering, School of Mechanical Engineering, University of Leeds, Leeds LS2 9JT, United Kingdom.

出版信息

J Bone Joint Surg Am. 2008 Aug;90 Suppl 3:111-7. doi: 10.2106/JBJS.H.00485.

Abstract

BACKGROUND

Clinical studies have indicated that the angular position of the acetabular cup may influence wear in metal-on-metal total hip bearings. A high cup angle in comparison to the anatomical position may lead to the head being constrained by the superior lateral surface and rim of the cup, thus potentially changing the location of the contact zone between the head and the cup. The aim of this study was to test the hypothesis that both a steep cup angle and a lateralized position of the head can increase head contact on the superior rim of the cup, with the consequence of increased wear.

METHODS

Hip-joint simulator studies of metal-on-metal bearings were undertaken with cup angles of 45 degrees and 55 degrees . The femoral head was either aligned to the center of the cup or placed in a position of microlateralization. Wear was measured gravimetrically over 5 million cycles.

RESULTS

A steep cup angle of 55 degrees showed significantly higher long-term steady-state wear than a standard cup angle of 45 degrees (p < 0.01). The difference was fivefold. Microlateralization of the head resulted in a fivefold increase in steady-state wear compared with a centralized head. The combination of a steep cup angle and a microlateralized head increased the steady-state wear rate by tenfold compared with a standard cup angle with a centralized head.

CONCLUSIONS

These studies support the hypothesis that both an increased cup angle and a lateral head position increase wear in metal-on-metal hip prostheses.

摘要

背景

临床研究表明,髋臼杯的角度位置可能会影响金属对金属全髋关节轴承的磨损。与解剖位置相比,髋臼杯角度过高可能导致股骨头受到髋臼杯上外侧表面和边缘的限制,从而可能改变股骨头与髋臼杯之间接触区域的位置。本研究的目的是检验以下假设:髋臼杯角度陡峭和股骨头侧向化都会增加股骨头与髋臼杯上缘的接触,从而增加磨损。

方法

对金属对金属轴承进行髋关节模拟器研究,髋臼杯角度分别为45度和55度。股骨头要么与髋臼杯中心对齐,要么置于微侧向化位置。通过重量法测量500万次循环后的磨损情况。

结果

55度的陡峭髋臼杯角度显示出的长期稳态磨损明显高于45度的标准髋臼杯角度(p < 0.01)。差异为五倍。与股骨头居中相比,股骨头微侧向化导致稳态磨损增加了五倍。与股骨头居中的标准髋臼杯角度相比,陡峭髋臼杯角度与微侧向化股骨头相结合使稳态磨损率增加了十倍。

结论

这些研究支持以下假设:髋臼杯角度增加和股骨头侧向化都会增加金属对金属髋关节假体的磨损。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验