Accuracy of quantitative reconstructions in SPECT/CT imaging.

作者信息

Shcherbinin S, Celler A, Belhocine T, Vanderwerf R, Driedger A

机构信息

Department of Radiology, University of British Columbia, 366-828 West 10th Avenue, Vancouver BC, V5Z 1L8, Canada.

出版信息

Phys Med Biol. 2008 Sep 7;53(17):4595-604. doi: 10.1088/0031-9155/53/17/009. Epub 2008 Aug 4.

Abstract

The goal of this study was to determine the quantitative accuracy of our OSEM-APDI reconstruction method based on SPECT/CT imaging for Tc-99m, In-111, I-123, and I-131 isotopes. Phantom studies were performed on a SPECT/low-dose multislice CT system (Infinia-Hawkeye-4 slice, GE Healthcare) using clinical acquisition protocols. Two radioactive sources were centrally and peripherally placed inside an anthropometric Thorax phantom filled with non-radioactive water. Corrections for attenuation, scatter, collimator blurring and collimator septal penetration were applied and their contribution to the overall accuracy of the reconstruction was evaluated. Reconstruction with the most comprehensive set of corrections resulted in activity estimation with error levels of 3-5% for all the isotopes.

摘要

本研究的目的是确定基于SPECT/CT成像的我们的OSEM-APDI重建方法对Tc-99m、In-111、I-123和I-131同位素的定量准确性。使用临床采集协议,在SPECT/低剂量多层CT系统(Infinia-Hawkeye-4层,GE医疗)上进行了体模研究。将两个放射性源分别置于中心和周边位置,放入充满非放射性水的人体胸部体模内。应用了衰减、散射、准直器模糊和准直器隔板穿透校正,并评估了它们对重建总体准确性的贡献。采用最全面校正集进行重建,所有同位素的活度估计误差水平为3-5%。