采用新型自适应降噪滤波器降低肝脏多排螺旋计算机断层扫描的辐射剂量

Radiation dose reduction in hepatic multidetector computed tomography with a novel adaptive noise reduction filter.

作者信息

Funama Yoshinori, Awai Kazuo, Miyazaki Osamu, Goto Taiga, Nakayama Yoshiharu, Shimamura Masamitchi, Hiraishi Kumiko, Hori Shinichi, Yamashita Yasuyuki

机构信息

Department of Radiological Sciences, School of Health Sciences, Kumamoto University, 4-24-1 Kuhonji, Kumamoto 862-0976, Japan.

出版信息

Radiat Med. 2008 Apr;26(3):171-7. doi: 10.1007/s11604-007-0202-y.

Abstract

PURPOSE

The aim of this study was to optimize a novel adaptive noise reduction filter based on patient body weight and to investigate its utility for improving the image quality of low-dose hepatic computed tomography (CT) scans.

MATERIALS AND METHODS

The tube current-time product was changed from 140 to 180 and from 60 to 100 mAs at standard- and low-dose CT, respectively, based on the body weights of 45 patients. Unenhanced and two-phase contrast-enhanced helical scans were obtained at the standard dose during the hepatic arterial and equilibrium phases. During the equilibrium phase, we obtained low-dose scans of the liver immediately after standard-dose scans. The low-dose CT images were postprocessed with the filter. Two radiologists visually evaluated artifacts in the liver parenchyma and its graininess, the sharpness of the liver contour, tumor conspicuity, homogeneity of the enhancement of the portal vein, and overall image quality.

RESULTS

There was no statistically significant difference between standard and filtered low-dose images with respect to artifacts in the liver, the graininess of the liver parenchyma, tumor conspicuity, homogeneity of enhancement of the portal vein, or overall image quality.

CONCLUSION

The adaptive noise reduction filter effectively reduced image noise. We confirmed the effectiveness of the filter by examining clinical hepatic images obtained at low-dose CT.

摘要

目的

本研究旨在基于患者体重优化一种新型自适应降噪滤波器,并研究其对改善低剂量肝脏计算机断层扫描(CT)图像质量的效用。

材料与方法

根据45例患者的体重,在标准剂量CT和低剂量CT时,管电流-时间乘积分别从140改变为180以及从60改变为100 mAs。在肝动脉期和平衡期以标准剂量进行未增强及两期对比增强螺旋扫描。在平衡期,在标准剂量扫描后立即进行肝脏的低剂量扫描。低剂量CT图像用该滤波器进行后处理。两名放射科医生对肝实质中的伪影及其颗粒度、肝脏轮廓的清晰度、肿瘤的显见度、门静脉强化的均匀性以及整体图像质量进行视觉评估。

结果

在肝脏伪影、肝实质颗粒度、肿瘤显见度、门静脉强化均匀性或整体图像质量方面,标准剂量图像与经滤波的低剂量图像之间无统计学显著差异。

结论

自适应降噪滤波器有效降低了图像噪声。通过检查低剂量CT获得的临床肝脏图像,我们证实了该滤波器的有效性。