Kinhikar R A, Master Z, Deshpande D D
Tata Memorial Hospital, Parel, Mumbai, India.
Australas Phys Eng Sci Med. 2008 Jun;31(2):139-45. doi: 10.1007/BF03178588.
Two-dimensional (2D) treatment verifications were performed for fifty patients planned and treated with helical tomotherapy (HT). The treatment verification consisted of an extended dose range (EDR2) film measurement as well as point dose measurements made with an A1SL ion chamber. The agreement between the calculated and the measured film dose distributions was evaluated with the gamma parameter calculated (3 mm and 3%). Good agreement was found between measured and calculated distributions with dose parameter registration using reference marks. The mean percent discrepancy for the point dose measurements was -0.4 (SD 1.4) for the high dose, low dose-gradient region. Thus the treatment verification results confirmed the safe delivery of the planned dose to the patient with helical tomotherapy.