[腰椎缓解矫形器治疗骶髂关节疼痛的生物力学效应证明]

[Proof of the biomechanical effect of a lumbar spine-relief orthosis for treatment of sacroiliac pain].

作者信息

Höfer S, Siemsen C H

机构信息

Biomechanik, HAW Hamburg, Fakultät LS, Buxtehude.

出版信息

Z Orthop Unfall. 2008 Jul-Aug;146(4):439-43. doi: 10.1055/s-2008-1038612.

Abstract

AIM

The use of a spine orthosis for the treatment of low back pain is an approved, but also a disputed therapy method. There have been numerous studies so far that attempt to examine the acceptance and the impact of this treatment method. But up to now there is no biomechanical assessment about the effective level of intervertebral disc decompression while wearing an orthosis. Therefore, the purpose of this study is to measure the reduction of the disc pressure while wearing an orthosis.

METHOD

The pressure between the sensor contact area and the lumbar region was measured with FSR sensors. Two sensors were fixed at the vertical rods of the plastic framework of the orthosis. The pressure sensors underwent a voltage change while a Tigges T-Flex-B orthosis was being worn. These voltage changes were recorded using the connected measurement amplifiers. Different series of measurements were conducted with 3 male test persons. The resulting pressures were recorded under the following test conditions: knees bent, in different flexion angles of the torso and lift of 10/20 kg weights. Back muscle, shear, compressive and resulting forces were computed afterwards with the generally valid biomechanical formulas.

RESULTS

The computations showed that the use of the Tigges T-Flex-B orthosis resulted in relief of the intervertebral disc L3-L4 of about approximately 96.91+/-4.68 N. The highest intervertebral disc pressure reduction was measured during the body-far lift of 20 kg weight.

CONCLUSION

For the first time relief of the lumbar spinal column was proven in the present study because the contact pressure was measured between the orthosis and the lumbal body surface.

摘要

目的

使用脊柱矫形器治疗腰痛是一种已获认可但也存在争议的治疗方法。迄今为止,已有众多研究试图检验这种治疗方法的接受度及其影响。但到目前为止,尚无关于佩戴矫形器时椎间盘减压有效程度的生物力学评估。因此,本研究的目的是测量佩戴矫形器时椎间盘压力的降低情况。

方法

使用FSR传感器测量传感器接触区域与腰椎区域之间的压力。两个传感器固定在矫形器塑料框架的垂直杆上。佩戴Tigges T-Flex-B矫形器时,压力传感器会发生电压变化。这些电压变化通过连接的测量放大器进行记录。对3名男性测试对象进行了不同系列的测量。在以下测试条件下记录所得压力:膝盖弯曲、躯干处于不同屈曲角度以及提起10/20千克重物。随后使用通用的生物力学公式计算背部肌肉力、剪切力、压缩力和合力。

结果

计算结果表明,使用Tigges T-Flex-B矫形器可使L3-L4椎间盘压力减轻约96.91±4.68牛。在提起20千克重物的身体远端提升过程中,测得的椎间盘压力降低幅度最大。

结论

本研究首次证明了腰椎得到了缓解,因为测量了矫形器与腰部体表之间的接触压力。