Barron Leon, O'Toole Martina, Diamond Dermot, Nesterenko Pavel N, Paull Brett
National Centre for Sensor Research, Dublin City University, Glasnevin, Dublin 9, Ireland.
J Chromatogr A. 2008 Dec 5;1213(1):31-6. doi: 10.1016/j.chroma.2008.08.049. Epub 2008 Aug 19.
The selectivity, retention and separation of transition metals on a short (2 mm x 50 mm) column packed with a poly-iminodiacetic acid functionalised polymer 10 microm resin (Dionex ProPac IMAC-10) are presented. This stationary phase, typically used for the separation of proteins, is composed of long chain poly-iminodiacetic acid groups grafted to a hydrophilic layer surrounding a 10 microm polymeric bead. Through the use of a combination of a multi-step pH and picolinic acid gradient, the separation of magnesium, iron, cobalt, cadmium, zinc, lead and copper was possible, followed by post-column reaction with 4-(2-pyridylazo) resorcinol (PAR) and absorbance detection at 510 nm using a novel and inexpensive optical detector, comprised of two light emitting diodes with one acting as a light source and the other as a detector. Column efficiency for selective transition metals was in excess of N=10,000, with the baseline separation of seven metal cations in <3 min possible under optimised conditions. Detection limits of between 5 and 81 microg/L were possible based upon a 50 microL injection volume.
介绍了过渡金属在填充有聚亚氨基二乙酸官能化聚合物10微米树脂(戴安ProPac IMAC - 10)的短柱(2毫米×50毫米)上的选择性、保留和分离情况。这种固定相通常用于蛋白质分离,由接枝到围绕10微米聚合物珠的亲水层上的长链聚亚氨基二乙酸基团组成。通过使用多步pH和吡啶甲酸梯度的组合,可以分离镁、铁、钴、镉、锌、铅和铜,然后与4 - (2 - 吡啶偶氮)间苯二酚(PAR)进行柱后反应,并使用一种新型且廉价的光学检测器在510纳米处进行吸光度检测,该检测器由两个发光二极管组成,一个作为光源,另一个作为检测器。选择性过渡金属的柱效超过N = 10000,在优化条件下,可在不到3分钟内实现七种金属阳离子的基线分离。基于50微升进样体积,检测限在5至81微克/升之间。