Faculty of Chemistry, Bu-Ali Sina University, Hamedan, 6517838695, Iran.
Research Center for Molecular Medicine, Hamadan University of Medical Sciences, Hamadan, Iran.
Sci Rep. 2023 Apr 17;13(1):6217. doi: 10.1038/s41598-023-33527-y.
This paper reports on developing a low cost but efficient paired emitter-detector diode (PEDD)-based photometer. The photometer consists of a white light-emitting diode (LED) as the emitter diode, an RGB LED as the detector diode, and a multimeter for recoding the signal. The developed PEDD-based photometer was utilized for the determination of liquid bleach adulteration in cow milk samples. N,N-Diethyl-p-phenylenediamine sulfate aqueous solution of pH 6 was used as a probe to monitor the presence of residual active chlorine in milk. The results showed that the developed method could be used to determine sodium hypochlorite in the concentration range of 0.5 to 20.0 ppm Cl with 0.14 and 0.46 ppm Cl limit of detection and limit of quantification, respectively. The intraday and interday precisions of the method at two concentration levels of 5.5 and 13.7 ppm Cl were 1.04% and 0.52%, and 1.81% and 1.02%, respectively. The recoveries of 114.2% and 106.9% were obtained for 5.5 and 13.7 ppm Cl concentrations levels, respectively. Real sample analyzes results showed that "maybe" liquid bleach adulteration in milk is the case for local distributors of raw milk.
本文报告了一种低成本但高效的基于配对发射极-探测器二极管(PEDD)的光度计的开发。该光度计由一个白色发光二极管(LED)作为发射二极管、一个 RGB LED 作为探测器二极管和一个用于记录信号的万用表组成。所开发的基于 PEDD 的光度计用于测定牛奶样品中液体漂白剂的掺假。pH 值为 6 的 N,N-二乙基对苯二胺硫酸盐水溶液被用作探针,以监测牛奶中残留的有效氯的存在。结果表明,该方法可用于测定浓度范围为 0.5 至 20.0 ppm Cl 的次氯酸钠,检测限和定量限分别为 0.14 和 0.46 ppm Cl。在 5.5 和 13.7 ppm Cl 两个浓度水平下,该方法的日内和日间精密度分别为 1.04%和 0.52%,1.81%和 1.02%。对于 5.5 和 13.7 ppm Cl 浓度水平,回收率分别为 114.2%和 106.9%。实际样品分析结果表明,当地生奶经销商的牛奶中“可能”存在液体漂白剂掺假。