Noninvasive temperature imaging using diffusion MRI.

作者信息

Delannoy J, Chen C N, Turner R, Levin R L, Le Bihan D

机构信息

Biomedical Engineering and Instrumentation Program, National Cancer Institute, National Institutes of Health, Bethesda, Maryland 20892.

出版信息

Magn Reson Med. 1991 Jun;19(2):333-9. doi: 10.1002/mrm.1910190224.

Abstract

Efficacy and safety considerations for cancer therapy with hyperthermia require accurate temperature measurements throughout the heated volume. We report the use of molecular diffusion, whose temperature dependence is well known. A dedicated hyperthermia applicator was built, combining a MRI gradient coil and a rf coil. Diffusion and derived temperature images were obtained with a 1 x 2 mm pixel size on a polyacrylamide gel phantom using a clinical 1.5-T whole body MRI system. Temperatures determined from these images using 1 cm2 regions of interest were found to be within 0.2 degrees C of those recorded from the thermocouples and fiber-optic probes placed inside the gel.