Sachs C, Rabouine P, Chaneac M, Kindermans C, Dechaux M
Department of Physiology, University of Paris V, Necker-Enfants Malades Medical School, France.
Scand J Clin Lab Invest. 1991 May;51(3):303-6. doi: 10.3109/00365519109091619.
We report modifications of ionized calcium as observed on specimens collected in a commercially available sampling device (Radiometer B-129). When the sampler is used as recommended, errors are non-significant around 1.25 mmol l-1. There is a +3% increase for low (0.75 mmol l-1) values and a -2% decrease for high (2.50 mmol l-1) values. Incomplete syringe-filling increases these errors.