Brenner Nina, Haapala Markus, Vuorensola Katariina, Kostiainen Risto
Division of Pharmaceutical Chemistry, Faculty of Pharmacy, P.O. Box 56, FI-00014 University of Helsinki, Finland.
Anal Chem. 2008 Nov 1;80(21):8334-9. doi: 10.1021/ac801406t. Epub 2008 Oct 7.
A simple method for direct coupling of gas chromatography (GC) with electrospray ionization mass spectrometry (ESI/MS) has been developed. The outlet of the GC capillary column was placed between the ESI needle and the atmospheric pressure ionization (API) source of a mass spectrometer. The ionization occurs via dissolution of neutral compounds into the charged ESI droplet followed by ion evaporation or via a gas-phase proton transfer reaction between a protonated solvent molecule and an analyte. The mass spectra of organic volatile compounds showed abundant protonated molecules with little fragmentation, being very similar to those produced by normal liquid ESI. The quantitative performance of the system was evaluated by determining the limit of detection (LOD), linearity ( r (2)), and repeatability (RSD). The GC-ESI/MS method was shown to be stable, providing high sensitivity and good quantitative performance.
已开发出一种将气相色谱(GC)与电喷雾电离质谱(ESI/MS)直接联用的简单方法。气相色谱毛细管柱的出口置于电喷雾针和质谱仪的大气压电离(API)源之间。电离通过中性化合物溶解于带电的电喷雾液滴中,随后进行离子蒸发,或者通过质子化溶剂分子与分析物之间的气相质子转移反应发生。有机挥发性化合物的质谱显示出大量质子化分子且几乎没有碎片化,与正常液体电喷雾产生的质谱非常相似。通过测定检测限(LOD)、线性度(r(2))和重复性(RSD)来评估该系统的定量性能。结果表明,GC-ESI/MS方法稳定,具有高灵敏度和良好的定量性能。