Daptomycin is the first approved member of the new class cyclic lipopeptide antibiotic drugs, effective against a broad spectrum of Gram-positive bacteria. Here we present an HPLC method with UV detection capable to obtain pharmacokinetic data of daptomycin in human plasma, exemplarily shown in a critically ill patient with acute renal failure undergoing extended daily dialysis. Sample preparation consists only of protein precipitation with methanol. Chromatographic separation was achieved on a Zorbax Eclipse XDB-C8 column and daptomycin was detected at 224 nm. The calibration function was linear over the range from 3.5 to 350 microg/ml. The relative standard deviations were <2% in the intra-day and <6% in the inter-day measurements. The accuracy was always better than 7%. Daptomycin was stable in aqueous solutions for 2 months frozen at-20 degrees C. However, in plasma frozen at -20 degrees C a loss of 25% in 1 month was observed.
达托霉素是新一类环脂肽抗生素药物中首个获批的成员,对多种革兰氏阳性菌有效。在此,我们展示一种具有紫外检测功能的高效液相色谱法,该方法能够获取人血浆中达托霉素的药代动力学数据,以一名接受延长每日透析的急性肾衰竭危重症患者为例进行说明。样品制备仅包括用甲醇进行蛋白沉淀。在Zorbax Eclipse XDB - C8柱上实现色谱分离,达托霉素在224 nm处检测。校准函数在3.5至350微克/毫升范围内呈线性。日内相对标准偏差<2%,日间测量相对标准偏差<6%。准确度始终优于7%。达托霉素在-20℃冷冻的水溶液中稳定2个月。然而,在-20℃冷冻的血浆中,1个月内观察到有25%的损失。