Wolthaus J W H, Sonke J J, van Herk M, Damen E M F
Department of Radiation Oncology, The Netherlands Cancer Institute-Antoni van L&uwenhoek Hospital, Plesmanlaan 121, 1066 CX Amsterdam, The Netherlands.
Med Phys. 2008 Sep;35(9):3998-4011. doi: 10.1118/1.2966347.
lower lobe lung tumors move with amplitudes of up to 2 cm due to respiration. To reduce respiration imaging artifacts in planning CT scans, 4D imaging techniques are used. Currently, we use a single (midventilation) frame of the 4D data set for clinical delineation of structures and radiotherapy planning. A single frame, however, often contains artifacts due to breathing irregularities, and is noisier than a conventional CT scan since the exposure per frame is lower. Moreover, the tumor may be displaced from the mean tumor position due to hysteresis. The aim of this work is to develop a framework for the acquisition of a good quality scan representing all scanned anatomy in the mean position by averaging transformed (deformed) CT frames, i.e., canceling out motion. A nonrigid registration method is necessary since motion varies over the lung.
4D and inspiration breath-hold (BH) CT scans were acquired for 13 patients. An iterative multiscale motion estimation technique was applied to the 4D CT scan, similar to optical flow but using image phase (gray-value transitions from bright to dark and vice versa) instead. From the (4D) deformation vector field (DVF) derived, the local mean position in the respiratory cycle was computed and the 4D DVF was modified to deform all structures of the original 4D CT scan to this mean position. A 3D midposition (MidP) CT scan was then obtained by (arithmetic or median) averaging of the deformed 4D CT scan. Image registration accuracy, tumor shape deviation with respect to the BH CT scan, and noise were determined to evaluate the image fidelity of the MidP CT scan and the performance of the technique.
Accuracy of the used deformable image registration method was comparable to established automated locally rigid registration and to manual landmark registration (average difference to both methods < 0.5 mm for all directions) for the tumor region. From visual assessment, the registration was good for the clearly visible features (e.g., tumor and diaphragm). The shape of the tumor, with respect to that of the BH CT scan, was better represented by the MidP reconstructions than any of the 4D CT frames (including MidV; reduction of "shape differences" was 66%). The MidP scans contained about one-third the noise of individual 4D CT scan frames.
We implemented an accurate method to estimate the motion of structures in a 4D CT scan. Subsequently, a novel method to create a midposition CT scan (time-weighted average of the anatomy) for treatment planning with reduced noise and artifacts was introduced. Tumor shape and position in the MidP CT scan represents that of the BH CT scan better than MidV CT scan and, therefore, was found to be appropriate for treatment planning.
由于呼吸作用,下叶肺部肿瘤移动幅度可达2厘米。为减少计划CT扫描中的呼吸成像伪影,采用了4D成像技术。目前,我们使用4D数据集的单个(呼气中期)帧进行结构的临床勾画和放射治疗计划。然而,单个帧常常因呼吸不规则而包含伪影,并且由于每帧的曝光较低,其噪声比传统CT扫描更大。此外,由于滞后现象,肿瘤可能会偏离平均肿瘤位置。这项工作的目的是开发一个框架,通过对变换(变形)后的CT帧进行平均,即消除运动,来获取代表平均位置所有扫描解剖结构的高质量扫描。由于肺部运动变化,需要一种非刚性配准方法。
对13名患者进行了4D和吸气屏气(BH)CT扫描。将一种迭代多尺度运动估计技术应用于4D CT扫描,类似于光流,但使用图像相位(从亮到暗以及反之的灰度值转变)代替。根据导出的(4D)变形向量场(DVF),计算呼吸周期中的局部平均位置,并修改4D DVF,将原始4D CT扫描的所有结构变形到该平均位置。然后通过对变形后的4D CT扫描进行(算术或中位数)平均获得3D中间位置(MidP)CT扫描。确定图像配准精度、相对于BH CT扫描的肿瘤形状偏差和噪声,以评估MidP CT扫描的图像保真度和该技术的性能。
对于肿瘤区域,所使用的可变形图像配准方法的精度与已建立的自动局部刚性配准和手动地标配准相当(所有方向与两种方法的平均差异<0.5毫米)。通过视觉评估,对于清晰可见的特征(如肿瘤和膈肌),配准效果良好。相对于BH CT扫描的肿瘤形状,MidP重建比任何4D CT帧(包括MidV)都能更好地呈现(“形状差异”减少了66%)。MidP扫描的噪声约为单个4D CT扫描帧噪声的三分之一。
我们实现了一种准确的方法来估计4D CT扫描中结构的运动。随后,引入了一种新颖的方法来创建用于治疗计划的中间位置CT扫描(解剖结构的时间加权平均值),其噪声和伪影减少。MidP CT扫描中的肿瘤形状和位置比MidV CT扫描更能代表BH CT扫描中的情况,因此被认为适用于治疗计划。