Suppr超能文献

连接质谱仪以在系统故障期间提供持续监测。

Linking mass spectrometers to provide continuing monitoring during system failure.

作者信息

Steinbrook R A, Elliott W R, Goldman D B, Philip J H

机构信息

Department of Anesthesia, Brigham and Women's Hospital, Boston, MA 02115.

出版信息

J Clin Monit. 1991 Jul;7(3):271-3. doi: 10.1007/BF01619273.

Abstract

Although multipatient monitoring with a time-shared mass spectrometer provides considerable cost advantages, failure of one component in a shared system can disrupt gas monitoring at all sites. We describe a simple method for linking two central mass spectrometer systems to provide continual monitoring during failure of one unit, without the need for time-consuming reconfiguration of individual patient sample line and display connections.

摘要

尽管使用分时质谱仪进行多患者监测具有显著的成本优势,但共享系统中的一个组件出现故障可能会干扰所有站点的气体监测。我们描述了一种简单的方法,可将两个中央质谱仪系统连接起来,以便在一个单元出现故障时提供持续监测,而无需对各个患者的样本管线和显示连接进行耗时的重新配置。

相似文献

2
One-way leak in mass spectrometer sampling system.质谱仪采样系统中的单向泄漏。
Anesthesiology. 1987 Oct;67(4):606-7. doi: 10.1097/00000542-198710000-00039.
4
Pitfalls with mass spectrometry in clinical anesthesia.临床麻醉中质谱分析的陷阱
Int J Clin Monit Comput. 1984;1(1):27-34. doi: 10.1007/BF01726659.