Surdo V, Dalla Pria P
Divisione di Ortopedia e Traumatologia dell'Ospedale di Mirano.
Ital J Orthop Traumatol. 1991 Mar;17(1):125-30.
Four cases of breakage of Küntscher intramedullary nails (AO model) are presented. All 4 nails broke after fracture healing had occurred and all 4 presented breakage of the stem in the same site. In these cases breakage of the nails is not due to instability of the fracture, but always derives from superficial micro-cracks, technologically inevitable, due to defects or irregularities in the metal that expand during load application.