Arndt Andreas, Nüsser Peter, Graichen Kurt, Müller Johannes, Lampe Bernhard
Berlin Heart GmbH, Berlin, Germany.
Artif Organs. 2008 Oct;32(10):761-71. doi: 10.1111/j.1525-1594.2008.00628.x.
A control strategy for rotary blood pumps meeting different user-selectable control objectives is proposed: maximum support with the highest feasible flow rate versus medium support with maximum ventricular washout and controlled opening of the aortic valve (AoV). A pulsatility index (PI) is calculated from the pressure difference, which is deduced from the axial thrust measured by the magnetic bearing of the pump. The gradient of PI with respect to pump speed (GPI) is estimated via online system identification. The outer loop of a cascaded controller regulates GPI to a reference value satisfying the selected control objective. The inner loop controls the PI to a reference value set by the outer loop. Adverse pumping states such as suction and regurgitation can be detected on the basis of the GPI estimates and corrected by the controller. A lumped-parameter computer model of the assisted circulation was used to simulate variations of ventricular contractility, pulmonary venous pressure, and aortic pressure. The performance of the outer control loop was demonstrated by transitions between the two control modes. Fast reaction of the inner loop was tested by stepwise reduction of venous return. For maximum support, a low PI was maintained without inducing ventricular collapse. For maximum washout, the pump worked at a high PI in the transition region between the opening and the permanently closed AoV. The cascaded control of GPI and PI is able to meet different control objectives and is worth testing in vitro and in vivo.
提出了一种用于旋转血泵的控制策略,该策略可满足不同的用户可选控制目标:以最高可行流速提供最大支持与以最大心室冲洗和主动脉瓣(AoV)的受控打开提供中等支持。脉动指数(PI)由压力差计算得出,该压力差由泵的磁轴承测量的轴向推力推导得出。通过在线系统辨识估计PI相对于泵速的梯度(GPI)。级联控制器的外环将GPI调节到满足所选控制目标的参考值。内环将PI控制到外环设定的参考值。基于GPI估计可以检测到诸如抽吸和反流等不良泵送状态,并由控制器进行校正。使用辅助循环的集总参数计算机模型来模拟心室收缩性、肺静脉压力和主动脉压力的变化。通过两种控制模式之间的转换展示了外部控制环的性能。通过逐步减少静脉回流来测试内环的快速反应。对于最大支持,维持低PI而不引起心室塌陷。对于最大冲洗,泵在AoV打开和永久关闭之间的过渡区域以高PI工作。GPI和PI的级联控制能够满足不同的控制目标,值得在体外和体内进行测试。