Burguera J L, Townshend A
Chemistry Department, Birmingham University, P.O. Box 363, Birmingham, B15 2TT, U.K.
Talanta. 1980 Apr;27(4):309-14. doi: 10.1016/0039-9140(80)80088-9.
Sulphide (0.1 ng/ml) is determined by measuring the chemiluminescence given by its oxidation in aqueous solution by hydrogen peroxide, catalysed by peroxidase. The detection limit is 0.05 ng/ml. Several other, rather less sensitive, systems are also described. Sulphite and thiocyanate, which depress the emission, are masked with formaldehyde.