Prokopowicz M, Luboch E, Namieśnik J, Biernat J F, Przyjazny A
Chemistry Faculty, Technical University of Gdańsk, 80-952 Gdańsk, Poland.
Talanta. 1997 Sep;44(9):1551-61. doi: 10.1016/s0039-9140(96)02177-7.
The utilization of a multicomponent gaseous standard mixture, containing CO and CO(2) and obtained by thermal decomposition of a so-called immobilized compound, for a single point calibration of a GC/FID system are described. The generation of such a mixture takes place as a result of thermal decomposition of a sample of chemically modified silica gel placed in a heated chamber of thermal desorber coupled with the device being calibrated via a catalytic methanizer. The mean amount of the analyte librated from unit mass of the gel was 0.71 mg g(-1) (RSD = 3%) for carbon monoxide and 0.86 mg g(-1) (RSD = 3%) for carbon dioxide.