Gholivand M B, Nozari N
Department of Chemistry, Faculty of Science, Razi University, Kermanshah, Iran.
Talanta. 2001 May 30;54(4):597-602. doi: 10.1016/s0039-9140(00)00671-8.
Poly(vinyl chloride) membrane electrode, that is highly selective and sensitive to Cu(II) ions, was developed by using 2,2'-dithiodianiline and dibutyl phthalate as carrier and plasticizer, respectively. The electrode exhibits good potentiometric response for Cu(II) over a wide concentration range (5.0x10(-2)-7.0x10(-7) mol l(-1)) with Nernstian slope of 30+/-1 mV per decade. The response time of the electrode is 10 s and it has been used for a period of one month and exhibits good selectivity towards Cu(2+) in comparison to alkali, alkaline earth, transition and heavy metal ions, with no interference caused by Pb(2+), Cd(2+) and Fe(+2) which are known to interfere with many other copper electrodes.