Centre de Phytopharmacie, Université de Perpignan-UMR CNRS 5054, 52, Avenue de Villeneuve, 66860 Perpignan Cedex, France; University of Bucharest, Faculty of Chemistry, Department of Analytical Chemistry, 90-92 Panduri, 76234, Bucharest, Romania.
Talanta. 2002 Apr 22;57(1):169-76. doi: 10.1016/s0039-9140(02)00017-6.
A screen-printed biosensor for the detection of pesticides in water miscible organic solvents is described based on the use of p-aminophenyl acetate as acetylcholinesterase substrate. The oxidation of p-aminophenol, product of the enzymatic reaction was monitored at 100 mV vs. Ag/AgCl screen-printed reference electrode. Miscible organic solvents as ethanol and acetonitrile were tested. The acetylcholinesterase (AChE) was immobilised on a screen-printed electrode surface by entrapment in a PVA-SbQ polymer and the catalytic activity of immobilised AChE was studied in the presence of different percentages of organic solvents in buffer solution. The sensor shows good characteristics when experiments were performed in concentrations of organic solvents below 10%. No significant differences were observed when working with 1 and 5% acetonitrile in the reaction media. Detection limits as low as 1.91x10(-8) M paraoxon and 1.24x10(-9) M chlorpyrifos ethyl oxon were obtained when experiments are carried out in 5% acetonitrile.
一种用于检测水混有机溶剂中农药的丝网印刷生物传感器,基于使用对氨基苯乙酸作为乙酰胆碱酯酶的底物。通过监测酶反应的产物对氨基酚在 100 mV 相对于 Ag/AgCl 丝网印刷参比电极的氧化来进行检测。测试了可混有机溶剂,如乙醇和乙腈。乙酰胆碱酯酶(AChE)通过包埋在 PVA-SbQ 聚合物中固定在丝网印刷电极表面上,并在缓冲溶液中存在不同百分比的有机溶剂的情况下研究了固定化 AChE 的催化活性。当在有机溶剂浓度低于 10%的情况下进行实验时,传感器表现出良好的特性。当在反应介质中使用 1%和 5%的乙腈时,没有观察到显著差异。当在 5%的乙腈中进行实验时,检测限低至 1.91x10(-8) M 对氧磷和 1.24x10(-9) M 毒死蜱乙基氧。