Tanikkul Sumalee, Jakmunee Jaroon, Lapanantnoppakhun Somchai, Rayanakorn Mongkon, Sooksamiti Ponlayuth, Synovec Robert E, Christian Gary D, Grudpan Kate
Department of Chemistry, Faculty of Science, Chiang Mai University, Chiang Mai 50200, Thailand.
Talanta. 2004 Dec 15;64(5):1241-6. doi: 10.1016/j.talanta.2004.05.060.
A flow injection (FI) in-valve-mini-column packed with Chelex-100 resin is proposed for on-line sample pretreatment for some metal ions, namely, Cd(II), Pb(II) and Zn(II), prior to simultaneous determination using ion chromatography (IC). A solution containing a mixture of the cations was first passed through the in-valve-mini-column, followed by on-line elution. The eluate was then flowed further to an injection valve and was injected into an ion chromatograph. Conditions of the system were optimized. A single standard calibration was possible. The recoveries of cations were found to be in the range of 95-105%. The developed method was applied to the accurate analysis of zinc ore samples.
提出了一种流动注射(FI)阀内微型柱,该微型柱填充有Chelex-100树脂,用于在离子色谱(IC)同时测定之前对某些金属离子,即镉(II)、铅(II)和锌(II)进行在线样品预处理。首先将含有阳离子混合物的溶液通过阀内微型柱,然后进行在线洗脱。洗脱液再进一步流至进样阀并注入离子色谱仪。对该系统的条件进行了优化。可以进行单点标准校准。发现阳离子的回收率在95%-105%范围内。所开发的方法应用于锌矿石样品的准确分析。