Udesen Jesper, Gran Fredrik, Hansen Kristoffer Lindskov, Jensen Jørgen Arendt, Thomsen Carsten, Nielsen Michael Bachmann
Dept. of Radiol., Rigshospitalet. Blegdamsvej, Copenhagen, Denmark.
IEEE Trans Ultrason Ferroelectr Freq Control. 2008 Aug;55(8):1729-43. doi: 10.1109/TUFFC.2008.858.
Conventional ultrasound methods for acquiring color images of blood velocity are limited by a relatively low frame-rate and are restricted to give velocity estimates along the ultrasound beam direction only. To circumvent these limitations, the method presented in this paper uses 3 techniques: 1) The ultrasound is not focused during the transmissions of the ultrasound signals; 2) A 13-bit Barker code is transmitted simultaneously from each transducer element; and 3) The 2-D vector velocity of the blood is estimated using 2-D cross-correlation. A parameter study was performed using the Field II program, and performance of the method was investigated when a virtual blood vessel was scanned by a linear array transducer. An improved parameter set for the method was identified from the parameter study, and a flow rig measurement was performed using the same improved setup as in the simulations. Finally, the common carotid artery of a healthy male was scanned with a scan sequence that satisfies the limits set by the Food and Drug Administration. Vector velocity images were obtained with a frame-rate of 100 Hz where 40 speckle images are used for each vector velocity image. It was found that the blood flow approximately followed the vessel wall, and that maximum velocity was approximately 1 m/s, which is a normal value for a healthy person. To further evaluate the method, the test person was scanned with magnetic resonance (MR) angiography. The volume flow derived from the MR scanning was compared with that from the ultrasound scanning. A deviation of 9% between the 2 volume flow estimates was found.
用于获取血流速度彩色图像的传统超声方法受到相对较低帧率的限制,并且仅限于仅沿超声束方向给出速度估计。为了克服这些限制,本文提出的方法使用了三种技术:1)在超声信号传输期间超声不聚焦;2)从每个换能器元件同时发送一个13位的巴克码;3)使用二维互相关估计血液的二维矢量速度。使用Field II程序进行了参数研究,并在使用线性阵列换能器扫描虚拟血管时研究了该方法的性能。从参数研究中确定了该方法的一组改进参数,并使用与模拟中相同的改进设置进行了流动试验台测量。最后,使用符合美国食品药品监督管理局设定限制的扫描序列对一名健康男性的颈总动脉进行扫描。以100 Hz的帧率获得矢量速度图像,其中每个矢量速度图像使用40个散斑图像。结果发现血流大致沿着血管壁流动,最大速度约为1 m/s,这对于健康人来说是正常数值。为了进一步评估该方法,使用磁共振(MR)血管造影对测试对象进行扫描。将MR扫描得出的体积流量与超声扫描得出的体积流量进行比较。发现两种体积流量估计值之间的偏差为9%。