Du Yigang, Ding Haiyan, He Le, Yiu Billy Y S, Deng Linsong, Yu Alfred C H, Zhu Lei
Shenzhen Mindray Bio-Medical Electronics Co., Ltd., Shenzhen 518057, China.
Department of Biomedical Engineering, Tsinghua University, Beijing 100084, China.
Diagnostics (Basel). 2022 Mar 11;12(3):690. doi: 10.3390/diagnostics12030690.
V Flow is commercially developed by high-frame-rate ultrasound vector flow imaging. Compared to conventional color Doppler, V Flow is angle-independent and is capable of measuring both the magnitude and the direction of blood flow velocities. This paper aims to investigate the differences between V Flow and pulsed wave Doppler (PW) relative to phase contrast magnetic resonance imaging (PC-MRI), for the quantitative measurements of blood flow in common carotid arteries (CCA) and, consequently, to evaluate the accuracy of the new technique, V Flow. Sixty-four CCAs were measured using V Flow, PW, and PC-MRI. The maximum velocities, time-averaged mean (TAMEAN) velocities, and volume flow were measured using different imaging technologies. The mean error with standard deviation (Std), the median of absolute errors, and the r-values between V Flow and PC-MRI results for the maximum velocity, the TAMEAN velocity, and the volume flow measurements are {9.40% ± 14.91%; 11.84%; 0.84}, {21.52% ± 14.46%; 19.28%; 0.86}, and {-2.80% ± 14.01%; 10.38%; 0.7}, respectively, and are {53.44% ± 29.68%; 49.79%; 0.74}, {27.83% ± 31.60%; 23.83; 0.71}, and {21.01% ± 29.64%; 25.48%; 0.34}, respectively, for those between PW and PC-MRI. The boxplot, linear regression and Bland-Altman plots were performed for each comparison, which illustrated that the results measured via V Flow rather than via PW agreed more closely with those measured via PC-MRI.
V 流是通过高帧率超声矢量流成像进行商业开发的。与传统彩色多普勒相比,V 流与角度无关,能够测量血流速度的大小和方向。本文旨在研究 V 流和脉冲波多普勒(PW)相对于相位对比磁共振成像(PC-MRI)在颈总动脉(CCA)血流定量测量方面的差异,从而评估新技术 V 流的准确性。使用 V 流、PW 和 PC-MRI 对 64 条颈总动脉进行了测量。使用不同的成像技术测量了最大速度、时间平均平均(TAMEAN)速度和体积流量。V 流与 PC-MRI 在最大速度、TAMEAN 速度和体积流量测量结果之间的平均误差与标准差(Std)、绝对误差中位数和 r 值分别为{9.40% ± 14.91%;11.84%;0.84}、{21.52% ± 14.46%;19.28%;0.86}和{-2.80% ± 14.01%;10.38%;0.7},而 PW 与 PC-MRI 之间的分别为{53.44% ± 29.68%;49.79%;0.74}、{27.83% ± 31.60%;23.83;0.71}和{21.01% ± 29.64%;25.48%;0.34}。对每次比较都进行了箱线图、线性回归和 Bland-Altman 图分析,结果表明通过 V 流而非 PW 测量的结果与通过 PC-MRI 测量的结果更为接近。