Dello Russo Antonio, Biddau Roberto, Pelargonio Gemma, Casella Michela, Frontera Antonio, Camporeale Antonella, Bisceglia Caterina, Forleo Giovanni B, Tondo Claudio, Zecchi Paolo, Bellocci Fulvio
Cardiac Arrhythmia and Heart Failure Research Center, Catholic University of Sacred Heart - St. Camillo-Forlanini Hospital, Rome, Italy.
J Interv Card Electrophysiol. 2009 Mar;24(2):147-50. doi: 10.1007/s10840-008-9326-y. Epub 2008 Nov 6.
We report the case of a 50 year-old woman referred for extraction of an infected biventricular implantable cardioverter defibrillator (ICD) who underwent device and leads extraction. Due to adhesions at the level of the subclavian-innominate vein angle, manual traction with locking stylet was ineffective. A new mechanical sheath provided with a stainless steel bladed tip, the Evolution Mechanical Dilator Sheath (Cook Medical), was successfully used to complete lead extraction. As far as we know, this is the first report on the use of this new tool; its safety should be ascertained by increasing the number of cases but its simple use and its low cost make the Evolution Mechanical Dilator Sheath a new interesting tool that can be added to the instruments available for lead extraction.