Suppr超能文献

调制超声参数对超声诱导溶栓的影响。

Effect of modulated ultrasound parameters on ultrasound-induced thrombolysis.

作者信息

Soltani Azita, Volz Kim R, Hansmann Doulas R

机构信息

Research and Development Department, EKOS Corporation, 11911 N Creek Parkway S, Bothell, WA 98011, USA.

出版信息

Phys Med Biol. 2008 Dec 7;53(23):6837-47. doi: 10.1088/0031-9155/53/23/012. Epub 2008 Nov 12.

Abstract

The potential of ultrasound to enhance enzyme-mediated thrombolysis by application of constant operating parameters (COP) has been widely demonstrated. In this study, the effect of ultrasound with modulated operating parameters (MOP) on enzyme-mediated thrombolysis was investigated. The MOP protocol was applied to an in vitro model of thrombolysis. The results were compared to a COP with the equivalent soft tissue thermal index (TIS) over the duration of ultrasound exposure of 30 min (p < 0.14). To explore potential differences in the mechanism responsible for ultrasound-induced thrombolysis, a perfusion model was used to measure changes in average fibrin pore size of clot before, after and during exposure to MOP and COP protocols and cavitational activity was monitored in real time for both protocols using a passive cavitation detection system. The relative lysis enhancement by each COP and MOP protocol compared to alteplase alone yielded values of 33.69 +/- 12.09% and 63.89 +/- 15.02% in a thrombolysis model, respectively (p < 0.007). Both COP and MOP protocols caused an equivalent significant increase in average clot pore size of 2.09 x 10(-2) +/- 0.01 microm and 1.99 x 10(-2) +/- 0.004 microm, respectively (p < 0.74). No signatures of inertial or stable cavitation were observed for either acoustic protocol. In conclusion, due to mechanisms other than cavitation, application of ultrasound with modulated operating parameters has the potential to significantly enhance the relative lysis enhancement compared to application of ultrasound with constant operating parameters.

摘要

超声通过应用恒定操作参数(COP)增强酶介导溶栓作用的潜力已得到广泛证实。在本研究中,研究了具有调制操作参数(MOP)的超声对酶介导溶栓作用的影响。将MOP方案应用于体外溶栓模型。将结果与在30分钟超声暴露期间具有等效软组织热指数(TIS)的COP进行比较(p<0.14)。为了探究超声诱导溶栓作用机制的潜在差异,使用灌注模型测量在暴露于MOP和COP方案之前、期间和之后血凝块平均纤维蛋白孔径的变化,并使用被动空化检测系统实时监测两种方案的空化活性。在溶栓模型中,与单独使用阿替普酶相比,每种COP和MOP方案的相对溶栓增强值分别为33.69±12.09%和63.89±15.02%(p<0.007)。COP和MOP方案均使血凝块平均孔径分别显著增加了2.09×10⁻²±0.01微米和1.99×10⁻²±0.004微米(p<0.74)。两种声学方案均未观察到惯性或稳定空化的信号。总之,由于空化以外的机制,与应用恒定操作参数的超声相比,应用具有调制操作参数的超声有可能显著提高相对溶栓增强效果。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验