Instrument fracture is a serious concern in endodontic treatment. Several studies have evaluated the influence of various factors on the fracture of nickel-titanium endodontic rotary instruments. The present study analyzes the influence of electrochemical polishing on flexural fatigue and torsional properties of K3 nickel-titanium endodontic rotary instruments. New files and polished files were tested for flexural fatigue and for resistance to fracture by twisting. The t test was used to compare the groups for number of cycles, angle of rotation, and maximum torque necessary to fracture. No statistical difference existed between these groups. Scanning electron microscopy analysis showed no significant differences caused by electrochemical polishing between these groups. These results suggest that electrochemical polishing has no influence on resistance to fracture of K3 rotary instruments.
器械折断是牙髓治疗中一个严重的问题。多项研究评估了各种因素对镍钛旋转牙髓器械折断的影响。本研究分析了电化学抛光对K3镍钛旋转牙髓器械弯曲疲劳和扭转性能的影响。对新锉和抛光后的锉进行了弯曲疲劳和扭转抗折断测试。采用t检验比较两组的循环次数、旋转角度和折断所需的最大扭矩。两组之间不存在统计学差异。扫描电子显微镜分析显示,电化学抛光在两组之间未造成显著差异。这些结果表明,电化学抛光对K3旋转器械的抗折断性没有影响。