Comparative analysis of lighted ureteral stents: lumination and tissue effects.

作者信息

Pedro Renato N, Kishore T A, Hinck Bryan D, Akornor Joseph W, Dickinson Leslie, Roychowdhury Monika, Anderson J Kyle, Monga Manoj

机构信息

Department of Urologic Surgery, University of Minnesota, Minneapolis, Minnesota 55455-0392, USA.

出版信息

J Endourol. 2008 Nov;22(11):2555-8. doi: 10.1089/end.2008.0278.

Abstract

PURPOSE

The objectives of the present study were to compare the luminescence of three types of ureteral illuminated stents and analyze their effects on urothelial histology.

MATERIALS AND METHODS

Three types of illuminating ureteral stents; the Cook single illuminating catheter, Cook double illuminating catheter and Stryker illumination system stent were laparoscopically placed in nine female white pigs (50 kg), under general anesthesia. After leaving the stents illuminated for 3 hours, during which time peritoneal insufflation was maintained at 18 mm Hg, the ureter was transected and the intraluminal temperature of the ureter was measured with a digital thermometer. The ureteral tissue was then harvested for histologic evaluation, and the animal was euthanized.

RESULTS

Statistical analysis confirmed that Stryker and Cook double illuminated stents were equally efficient in illuminating the ureter (P 0.46) whereas, the Cook single stent was significantly superior (P = 0.000004). There was no significant difference in mean intraluminal temperatures between the Cook single (95.2 degrees F), Cook double (92.3 degrees F), and Stryker (95.1 degrees F) stents. When compared with the intraluminal temperature of control unstented ureters, no significant increase was noted with the Cook single (P = 0.85), Cook double (P = 0.57), or Stryker (P = 0.82). Histologic analysis did not show any evidence of thermal injury to the urothelium or any remarkable alteration in the ureteral mucosa.

CONCLUSION

The Cook single illuminated stent presented the highest luminescence. All three devices did not cause any remarkable injury to the urothelium after 3 hours of exposure.

摘要

目的

本研究的目的是比较三种类型的输尿管照明支架的发光情况,并分析它们对尿路上皮组织学的影响。

材料与方法

三种类型的照明输尿管支架;Cook单照明导管、Cook双照明导管和Stryker照明系统支架,在全身麻醉下通过腹腔镜放置在9只雌性白色猪(50千克)体内。将支架照明3小时,在此期间腹腔内压力维持在18毫米汞柱,然后切断输尿管,用数字温度计测量输尿管腔内温度。接着采集输尿管组织进行组织学评估,随后对动物实施安乐死。

结果

统计分析证实,Stryker和Cook双照明支架在照亮输尿管方面效率相同(P>0.46),而Cook单支架明显更优(P = 0.000004)。Cook单支架(95.2华氏度)、Cook双支架(92.3华氏度)和Stryker支架(95.1华氏度)的平均腔内温度之间没有显著差异。与未放置支架的对照输尿管腔内温度相比,Cook单支架(P = 0.85)、Cook双支架(P = 0.57)或Stryker支架(P = 0.82)均未发现显著升高。组织学分析未显示尿路上皮有热损伤的任何证据,输尿管黏膜也没有任何明显改变。

结论

Cook单照明支架的发光度最高。所有三种装置在暴露3小时后均未对尿路上皮造成任何明显损伤。