Department of Urologic Sciences, University of British Columbia, Vancouver, British Columbia, Canada.
J Endourol. 2011 Jun;25(6):1051-6. doi: 10.1089/end.2010.0722. Epub 2011 May 13.
To determine whether the overall ureteral flow through an obstructed ureter using the 3F MicroStent™ that uses a novel film occlusion anchoring mechanism is comparable to the flow using a conventional 3F and 4.7F Double-J stent.
An in vitro silicone ureter model and an ex vivo porcine urinary model (kidney and ureter) were used to measure the overall flow through obstructed and unobstructed ureters with either a 3F Double-J stent (Cook), 3F MicroStent (PercSys), or 4.7F Double-J stent (Cook). Mean flow rates were compared with descriptive statistics.
Mean flow rates through the obstructed silicone ureter (12-mm stone) for the 3F MicroStent, 3F Double-J stent, and 4.7F Double-J stent were 326.7±13.3 mL/min, 283.3±19.2 mL/min, and 356.7±14.1 mL/min, respectively. In the obstructed ex vivo porcine ureter model, the flow as a percentage of free flow was 60%, 53%, and 50 %, respectively. In both ureteral models, flow rates of the 3F MicroStent and 4.7F Double-J stents were not statistically different.
The 3F MicroStent demonstrated drainage equivalent to a 4.7F Double-J stent, in both in vitro silicone and ex vivo porcine obstructed urinary models. We have demonstrated the crucial first step that this 3F stent, using a novel film occlusion anchoring mechanism, has equivalent, if not slightly improved, drainage rates when compared with its larger counterpart.
确定使用新型薄膜闭塞锚固机制的 3F MicroStent™ 通过阻塞输尿管的整体输尿管流量是否与使用传统的 3F 和 4.7F Double-J 支架的流量相当。
使用体外硅树脂输尿管模型和离体猪泌尿模型(肾脏和输尿管)来测量阻塞和未阻塞输尿管的整体流量,方法是使用 3F Double-J 支架(Cook)、3F MicroStent(PercSys)或 4.7F Double-J 支架(Cook)。使用描述性统计方法比较平均流速。
3F MicroStent、3F Double-J 支架和 4.7F Double-J 支架通过阻塞的硅树脂输尿管(12mm 结石)的平均流速分别为 326.7±13.3 mL/min、283.3±19.2 mL/min 和 356.7±14.1 mL/min。在阻塞的离体猪输尿管模型中,自由流动的百分比分别为 60%、53%和 50%。在这两种输尿管模型中,3F MicroStent 和 4.7F Double-J 支架的流速均无统计学差异。
在体外硅树脂和离体猪阻塞性泌尿模型中,3F MicroStent 表现出与 4.7F Double-J 支架相当的引流效果。我们已经证明了这一关键的第一步,即这种 3F 支架,使用新型薄膜闭塞锚固机制,与较大的支架相比,具有相等的,如果不是略有改善的,引流率。