Etsadashvili Kakhaber, Hintringer Florian, Stühlinger Markus, Dichtl Wolfgang, Spuller Karin, Antretter Herwig, Hangler Herbert, Pachinger Otmar, Roithinger Franz Xaver, Berger Thomas
Clinical Department of Cardiology, Medical University Innsbruck Anichstrasse, Innsbruck, Austria.
Europace. 2009 Feb;11(2):200-5. doi: 10.1093/europace/eun328. Epub 2008 Dec 4.
The long-term effects of high impedance vs. standard impedance pacing leads on actual generator longevity were studied.
In 40 patients (21 females, age 73 +/- 13 years) with standard dual-chamber pacemaker indication, bipolar standard impedance ventricular leads and high-impedance leads were implanted in a randomized fashion. Identical pacemaker generators and atrial pacing leads were implanted in all patients. Patients were observed during a mean follow-up of 89.8 +/- 8.8 months before pacemaker replacement. Initially, the patients who received the high-impedance leads had a lower current drain as compared with standard pacing impedance leads, and the estimated pacemaker longevity was significantly prolonged, too. But this pattern disappeared after 6 years of follow-up, and finally the actual pacemaker generators' replacement time was 86.7 +/- 6.8 months in standard impedance lead group vs. 91.2 +/- 10.3 months in high-impedance lead group (P = 0.17).
Implantation of high pacing impedance leads for ventricular stimulation does not result in a benefit with respect to pacemaker longevity as compared with standard impedance leads.
研究高阻抗与标准阻抗起搏导线对实际起搏器使用寿命的长期影响。
在40例(21例女性,年龄73±13岁)有标准双腔起搏器适应证的患者中,以随机方式植入双极标准阻抗心室导线和高阻抗导线。所有患者均植入相同的起搏器发生器和心房起搏导线。在更换起搏器前,对患者进行了平均89.8±8.8个月的随访。最初,与标准起搏阻抗导线相比,接受高阻抗导线的患者电流消耗较低,估计的起搏器使用寿命也显著延长。但随访6年后这种模式消失,最终标准阻抗导线组实际起搏器发生器的更换时间为86.7±6.8个月,高阻抗导线组为91.2±10.3个月(P=0.17)。
与标准阻抗导线相比,植入高起搏阻抗导线进行心室刺激对起搏器使用寿命无益处。