Suppr超能文献

[Low stimulation impedance in pacemaker patients with cardiac diseases. The Thera Pacemaker Study Group].

作者信息

Schuchert A, van Langen H, Michels K, Meinertz T

机构信息

Abteilung für Kardiologie, Universitäts-Krankenhaus Eppendorf Hamburg.

出版信息

Dtsch Med Wochenschr. 1996 Aug 23;121(34-35):1046-9. doi: 10.1055/s-2008-1043105.

Abstract

BACKGROUND AND AIM OF THE STUDY

As increasing impedance of the pacing impulse decreases pacing current, high pacing impedance lowers pacing current and subsequently prolongs pacemaker longevity. The present assumption of a standard 500 ohms impedance disregards interindividual differences. In the present study, the interindividual range of pacing impedance was assessed at implantation. Additionally clinical parameters of patients with low pacing impedance were compared to those of patients with high values.

PATIENTS AND METHODS

The same lead and the same pacemaker (CapSure VDD; Thera VDD; Medtronic) were implanted in 126 patients (79 men, 47 women, mean age 65 +/- 16 years). Patients with an impedance below or similar to the median were compared to patients with an impedance above the median in respect to clinical (age, gender, arterial hypertension, cardiac diseases such as coronary heart disease, valvular heart disease, cardiomyopathy) and electrical parameters (pacing threshold at 0.5 ms pulse duration, R-wave amplitude).

RESULTS

Mean impedance was 687 +/- 121 ohms with a range from 333 to 1040 ohms and a median of 670 ohms. 64 patients had an impedance < or = 670 ohms and 62 patients > 670 ohms. Mean age, gender and the occurrence of arterial hypertension were similar in the two groups. Cardiac diseases occurred significantly (P < 0.01) more often in patients with lower (n = 23) than with higher impedance (n = 9). Intraoperative pacing thresholds and R-wave amplitudes were not different between the two groups.

CONCLUSIONS

The range of pacing impedance was 707 ohms for the studied lead. Under consideration of permanent pacing with 2.5 V pulse amplitude and 0.5 ms pulse duration this would vary expected longevity of the implanted pacemaker from 7.8 to 9.3 years. Patients with lower impedance more often had cardiac diseases than patients with higher impedance.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验