Suppr超能文献

单极髋关节间隔器结构的力学评估

Mechanical evaluation of unipolar hip spacer constructs.

作者信息

Kummer Frederick J, Strauss Eric, Wright Kevin, Kubiak Erik N, Di Cesare Paul E

机构信息

Department of Orthopaedic Surgery, New York University/Hospital for Joint Diseases, New York, NY 10003, USA.

出版信息

Am J Orthop (Belle Mead NJ). 2008 Oct;37(10):517-8.

Abstract

The strengths of 3 hip spacer constructs--Steinmann pins, a short intramedullary nail (both cement-incorporated), and a Charnley prosthesis--were determined and compared with the strength of a commercially available hip spacer. The hip prosthesis construct was more than twice as strong as the other 2 constructs and was equivalent in strength to the commercial spacer. For spacer applications in which limited weight-bearing is anticipated, the hip prosthesis construct appears more efficacious, but its pros and cons should be compared with those of the commercial product.

摘要

对三种髋关节间隔器结构——斯氏针、短髓内钉(均含骨水泥)和查恩利假体——的强度进行了测定,并与一种市售髋关节间隔器的强度进行了比较。髋关节假体结构的强度是其他两种结构的两倍多,且与市售间隔器的强度相当。对于预期负重有限的间隔器应用,髋关节假体结构似乎更有效,但应将其优缺点与市售产品的优缺点进行比较。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验