主动脉瓣置换术中主动脉内滤器装置捕获颗粒性栓子

Particulate emboli capture by an intra-aortic filter device during aortic valve replacement.

作者信息

Leite Filho Osanan Amorim, Brandão Carlos Manuel de Almeida, Pomerantzeff Pablo Maria Alberto, Guedes Marco Antônio Vieira, Higuchi Maria de Lourdes, Stolf Noedir Antônio Groppo

机构信息

Instituto do Coração, HC, FMUSP, São Paulo, Brasil..

出版信息

Rev Bras Cir Cardiovasc. 2008 Jul-Sep;23(3):431-5. doi: 10.1590/s0102-76382008000300026.

Abstract

OBJECTIVE

This study aims to analyze the embolic activity in patients with calcified aortic stenosis who underwent aortic valve replacement using intra-aortic filtration with an EMBOL-X System device (Edwards Lifesciences Inc., Mountain View, CA, USA).

METHODS

From January 2007 to July 2007, 13 consecutive patients with calcified aortic stenosis, who underwent isolated aortic valve replacement using intra-aortic filtration by an EMBOL-X System for 5 minutes after aortic clamp release, were evaluated. Mean patient age was 63.7 years (range 34 to 79 years) and 61.5% were female. The mean bypass time was 60.2 +/- 7.5 minutes (range 45 to 72 minutes) and the mean cross-clamp time was 50 +/- 7.5 minutes (range 35 to 63 minutes). Following removal, each filter was fixed in formalin and analyzed macroscopically with the captured fragments being counted. Histological examinations of the captured material were performed.

RESULTS

There were no strokes or gross neurological events. There were no cases of postoperative renal failure. No deaths were reported during hospitalization. Particulate emboli were found in five (38.5%) of the filters. On histological analysis of the particulate emboli captured, two (40%) contained fibrin, two (40%) presented conjunctive tissue, one (20%) contained red blood cells and in one it was not possible to determine the nature of the particulates captured.

CONCLUSION

The EMBOL-X System device was effective in particulate emboli capture in aortic valve replacement surgery of patients with calcified aortic stenosis.

摘要

目的

本研究旨在分析钙化性主动脉瓣狭窄患者在使用EMBOL-X系统装置(美国加利福尼亚州山景城爱德华兹生命科学公司)进行主动脉内过滤的主动脉瓣置换术中的栓塞活性。

方法

对2007年1月至2007年7月期间连续13例钙化性主动脉瓣狭窄患者进行评估,这些患者在主动脉阻断钳松开后使用EMBOL-X系统进行主动脉内过滤5分钟,接受单纯主动脉瓣置换术。患者平均年龄为63.7岁(范围34至79岁),女性占61.5%。平均体外循环时间为60.2±7.5分钟(范围45至72分钟),平均阻断时间为50±7.5分钟(范围35至63分钟)。取出后,每个过滤器用福尔马林固定并进行宏观分析,对捕获的碎片进行计数。对捕获的物质进行组织学检查。

结果

未发生中风或严重神经系统事件。无术后肾衰竭病例。住院期间无死亡报告。在5个(38.5%)过滤器中发现了颗粒状栓子。对捕获的颗粒状栓子进行组织学分析,其中2个(40%)含有纤维蛋白,2个(40%)呈现结缔组织,1个(20%)含有红细胞,1个无法确定捕获颗粒的性质。

结论

EMBOL-X系统装置在钙化性主动脉瓣狭窄患者的主动脉瓣置换手术中对颗粒状栓子的捕获有效。