Performance of bone cements: are current preclinical specifications adequate?

作者信息

Nottrott Markus, Mølster Anders O, Moldestad Irene O, Walsh William R, Gjerdet Nils R

机构信息

Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen, Norway.

出版信息

Acta Orthop. 2008 Dec;79(6):826-31. doi: 10.1080/17453670810016920.

Abstract

BACKGROUND

Current specifications (standards) for preclinical testing of bone cements (ISO 5833: 2002, ASTM F451-99a) require simple mechanical testing after ageing for 24 h under dry conditions at 23 degrees C. Some bone cements have fulfilled the requirements in the specifications, and yet had inferior clinical results. Clinically, bone cements are subjected to complex loading patterns in a moist or wet environment at 37 degrees C. Thus, both the validity and the robustness of current standard testing protocols can be questioned.

METHODS

We examined the influence of temperature and storage medium on the properties of bone cement. We also compared the results of storage and testing under standard conditions of 23 degrees C in dry air, with the results obtained at 37 degrees C in water or plasma.

RESULTS

The dry specimens showed an increase in strength and elastic modulus with time, while the values of the wet ones decreased. There was no difference between specimens stored in water or in plasma. Ultimate compressive strength of dry specimens after 24 h was 1.16 times higher than that of the ones stored wet, increasing to 1.34 times after 1 month, and 1.46 times after 6 months (p<0.001 for all comparisons).

INTERPRETATION

Testing under dry conditions-as required in current standards-always gave higher values for mechanical properties than did storage and testing under more physiological conditions. The sensitivity of test values to different environments implies that testing conditions for bone cements should be scrutinized in order to develop more relevant testing protocols that reflect the in vivo environment more closely.

摘要

背景

目前骨水泥临床前测试的规范(标准)(ISO 5833: 2002,ASTM F451 - 99a)要求在23摄氏度干燥条件下老化24小时后进行简单的机械测试。一些骨水泥满足了规范中的要求,但临床结果却较差。临床上,骨水泥在37摄氏度的潮湿或湿润环境中承受复杂的加载模式。因此,当前标准测试方案的有效性和稳健性都可能受到质疑。

方法

我们研究了温度和储存介质对骨水泥性能的影响。我们还比较了在23摄氏度干燥空气中的标准条件下储存和测试的结果,与在37摄氏度水中或血浆中获得的结果。

结果

干燥的标本强度和弹性模量随时间增加,而湿润标本的值下降。储存在水中或血浆中的标本之间没有差异。干燥标本24小时后的极限抗压强度比湿润储存的标本高1.16倍,1个月后增加到1.34倍,6个月后增加到1.46倍(所有比较p<0.001)。

解读

按照当前标准要求在干燥条件下进行测试,其机械性能值总是高于在更接近生理条件下的储存和测试。测试值对不同环境的敏感性意味着,应仔细审查骨水泥的测试条件,以便制定更相关的测试方案,更紧密地反映体内环境。