Impact of a prolonged delivery inflation time for optimal drug-eluting stent expansion.

作者信息

Kawasaki Tomohiro, Koga Hisashi, Serikawa Takeshi, Orita Yoshiya, Ikeda Shinsuke, Mito Takahiro, Gotou Yoshitaka, Shintani Yoshiaki, Tanaka Atsushi, Tanaka Hidenori, Fukuyama Takaya, Koga Nobuhiko

机构信息

Cardiovascular Center, Shin-Koga Hospital, Tenjin-cho, Kurume, Japan.

出版信息

Catheter Cardiovasc Interv. 2009 Feb 1;73(2):205-11. doi: 10.1002/ccd.21813.

Abstract

PURPOSE

We examined the importance of prolonged inflation time for optimal sirolimus-eluting stent (SES) or paclitaxel-eluting stent (PES) expansion.

METHODS

Eighty-one de novo lesions deployed single SES or PES between April 2007 and March 2008 were divided into four groups; group 1: 21 SES deployed at 20 atm x 60 sec, group 2: 20 SES deployed with 2-step inflation at 20 atm x 60 sec following 20 atm x 20 sec, group 3: 20 PES deployed same as group 1, group 4: 20 PES deployed same as group 2. The minimal lumen diameter (MLD) and stent expansion ratio (SER; stent cross- sectional area at lesion/balloon cross-sectional area which was calculated according to the compliance chart at the same atmosphere as stent deployment) were compared between group 1 and group 2 in SES, between group 3 and group 4 in PES.

RESULTS

The MLD of post 60 sec was significantly higher than that of post 20 sec (2.84 +/- 0.28 mm in group 1, 2.76 +/- 0.33 mm in group 2 vs. 2.54 +/- 0.33 mm in group 2; P = 0.003, 0.045, respectively and 2.94 +/- 0.28 mm in group 3, 3.00 +/- 0.34 mm in group 4 vs. 2.69 +/- 0.35 mm in group 4; P = 0.022, 0.007, respectively). The SER of post 60 sec was significantly higher than that of post 20 sec (79.3% +/- 8.5% in group 1, 80.8% +/- 7.8% in group 2 vs. 71.1% +/- 10.2% in group 2; P = 0.014, 0.011, respectively and 81.1% +/- 7.9% in group 3, 84.3% +/- 9.9% in group 4 vs. 72.6% +/- 10.5% in group 4, P = 0.011, 0.001, respectively).

CONCLUSION

The prolonged delivery inflation for 60 sec may result in a more optimal stent expansion. It is therefore considered to be a useful method for deploying drug-eluting stent.