Ressler Noel, Rider Alan R, Kunselman Allen R, Richardson J Scott, Dasse Kurt A, Wang Shigang, Undar Akif
Pediatric Cardiac Research Laboratory, Department of Health Evaluation Sciences, Penn State Milton S. Hershey Medical Center, Penn State College Medicine, Penn State Hershey Children's Hospital, Hershey, Pennsylvania 17033-0850, USA.
ASAIO J. 2009 Jan-Feb;55(1):106-10. doi: 10.1097/MAT.0b013e3181904232.
The hemodynamic comparison of the Jostra HL-20 and the Levitronix PediVAS blood pumps is the focus this study, where pressure-flow waveforms and hemodynamic energy values are analyzed in the confines of a pediatric cardiopulmonary bypass circuit.The pseudo pediatric patient was perfused with flow rates between 500 and 900 ml/min (100 ml/min increments) under pulsatile and nonpulsatile mode. The Levitronix continuous flow pump utilized a customized controller to engage in pulsatile perfusion with equivalent pulse settings to the Jostra HL-20 roller pump. Hemodynamic measurements and waveforms were recorded at the precannula location, while the mean arterial pressure was maintained at 40 mm Hg for each test. Glycerin water was used as the blood analog circuit perfusate. At each flow rate 24 trials were conducted yielding a total of 120 experiments (n=60 pulsatile and n=60 nonpulsatile).Under nonpulsatile perfusion the Jostra roller pump produced small values for surplus hemodynamic energy (SHE) due to its inherent pulsatility, while the Levitronix produced values of essentially zero for SHE. When switching to pulsatile perfusion, the SHE levels for both the Jostra and Levitronix pump made considerable increases. In comparing the two pumps under pulsatile perfusion, the Levitronix PediVAS produced significantly more surplus and total hemodynamic energy than did the Jostra roller pump each pump flow rate.The study suggests that the Levitronix PediVAS centrifugal pump has the capability of achieving quality pulsatile waveforms and delivering more SHE to the pseudo patient than the Jostra HL-20 roller pump. Further studies are warranted to investigate the Levitronix under bovine blood studies and with various pulsatile settings.
本研究的重点是比较Jostra HL - 20和Levitronix PediVAS血泵的血流动力学,在小儿体外循环回路范围内分析压力 - 流量波形和血流动力学能量值。在搏动和非搏动模式下,以500至900毫升/分钟(每次增加100毫升/分钟)的流速对模拟小儿患者进行灌注。Levitronix连续流泵使用定制控制器,以与Jostra HL - 20滚压泵等效的脉冲设置进行搏动灌注。在插管前位置记录血流动力学测量值和波形,每次测试时平均动脉压维持在40毫米汞柱。甘油水用作血液模拟回路灌注液。在每个流速下进行24次试验,总共进行了120次实验(n = 60次搏动和n = 60次非搏动)。在非搏动灌注下,Jostra滚压泵因其固有的搏动性产生的剩余血流动力学能量(SHE)值较小,而Levitronix产生的SHE值基本为零。当切换到搏动灌注时,Jostra和Levitronix泵的SHE水平都有显著增加。在搏动灌注下比较这两种泵时,在每个泵流速下,Levitronix PediVAS产生的剩余和总血流动力学能量明显多于Jostra滚压泵。该研究表明,Levitronix PediVAS离心泵能够实现高质量的搏动波形,并比Jostra HL - 20滚压泵向模拟患者输送更多的SHE。有必要进一步研究Levitronix在牛血研究和各种搏动设置下的情况。