Prasad S C, Bassano D A
Department of Radiology, SUNY, Syracuse 13210.
Med Dosim. 1991 Sep;16(3):159-61. doi: 10.1016/0958-3947(91)90127-n.
Dose calculations on a commercial treatment planning computer based on the storage of profile data along a principal axis for Clinac-4 4MV x-ray can lead to significant error in calculated dose in the corners of a large field. These errors are due to the unique nature of the lead flattening filter on Cl-4. A method is suggested to remedy this problem by storing profile data along a diagonal of a large field.