我们可以使用磷光板多少次?一项初步研究。

How many times can we use a phosphor plate? A preliminary study.

作者信息

Ergün S, Güneri P, Ilgüy D, Ilgüy M, Boyacioglu H

机构信息

Department of Oral Diagnosis and Radiology, Ege University, Bornova 35100, Izmir, Turkey.

出版信息

Dentomaxillofac Radiol. 2009 Jan;38(1):42-7. doi: 10.1259/dmfr/61622880.

Abstract

OBJECTIVES

Digital radiography has become a useful tool in daily dental practice due to the advances in imaging technologies. Charge coupled devices (CCDs) and photostimulable phosphor plates (PSPs) are currently in use for dental imaging; however, the longevity of PSPs in dental practice is not yet established. The aim of this study was to determine the service life of PSPs in a clinical setting.

METHODS

Five unused PSPs were exposed with a conventional X-ray device and converted into digital images with Digora Optime (Soredex, Milwaukee, WI). These were recorded as the baseline images. Subsequent digital images of the plates were obtained after 20, 40, 60, 80, 100, 120, 140, 160, 180 and 200 exposures. All radiographic images were subtracted from the first digital image (baseline) and the mean grey values (MGVs) of the subtracted images were established using software. The data were grouped in 3 classes according to the number of exposures (20-80; 100-140; 160-200), and were analysed using variance analysis and chi(2) tests.

RESULTS

The MGVs of the subtracted images varied between 126.25 and 127.59, and the difference was not significant among the groups (P = 0.11). However, the differences between the MGVs of the plates on each exposure settings were significantly different than those of the baseline image (P < 0.05).

CONCLUSIONS

The findings of this study revealed that even though a slight deterioration occurred after the first exposure, each plate can be used up to 200 times. Further studies are required to reach a more concrete conclusion.

摘要

目的

由于成像技术的进步,数字放射摄影已成为日常牙科实践中的一种有用工具。电荷耦合器件(CCD)和光激励荧光板(PSP)目前用于牙科成像;然而,PSP在牙科实践中的使用寿命尚未确定。本研究的目的是确定PSP在临床环境中的使用寿命。

方法

使用传统X射线设备对五块未使用的PSP进行曝光,并用Digora Optime(Soredex,密尔沃基,威斯康星州)将其转换为数字图像。这些被记录为基线图像。在20、40、60、80、100、120、140、160、180和200次曝光后获得板的后续数字图像。从第一张数字图像(基线)中减去所有射线照片图像,并使用软件确定减去图像的平均灰度值(MGV)。根据曝光次数(20 - 80;100 - 140;160 - 200)将数据分为3组,并使用方差分析和卡方检验进行分析。

结果

减去图像的MGV在126.25至127.59之间变化,各组之间差异不显著(P = 0.11)。然而,在每个曝光设置下板的MGV与基线图像的MGV之间的差异显著不同(P < 0.05)。

结论

本研究结果表明,尽管第一次曝光后会出现轻微劣化,但每块板最多可使用200次。需要进一步研究以得出更具体的结论。