Böck J C, Sander B, Frank J, Schörner W
Strahlenklinik und Poliklinik, Universitätsklinikum Rudolf Virchow, Freie Universität Berlin.
Rofo. 1991 Sep;155(3):267-71. doi: 10.1055/s-2008-1033259.
We studied first-pass MRI contrast dilution to compute flow and volume of distribution in a realistic flow phantom. Pulsatile flow was provided by a one-chamber artificial heart. Physiological stroke volume, rate, pressure, and flow were adjustable. An elastic tube with dimensions similar to that of the human aorta was imaged at a rate of 2.4 Hz. After contrast injection, an initial increase in signal intensity was followed by a decrease. Signal-intensity time plots demonstrated slightly skewed curves as expected from dispersion theory. After calibration at different gadolinium-DTPA concentrations, signal intensities were converted into true gadolinium concentrations, and flow was calculated from the concentration-time curves. Flow was varied between 2.5 and 10.0 l/min and a significant correlation was found between the MRI estimate and true flow. Volume of distribution between injection and detection site was reliably estimated. This study demonstrates rapid 2-D imaging of a paramagnetic contrast bolus in a realistic flow phantom. Reliable estimates of flow and volume are obtained.