锥形束计算机断层扫描对下颌管区域的测量:一项尸体研究。

Measurements of mandibular canal region obtained by cone-beam computed tomography: a cadaveric study.

作者信息

Kamburoğlu Kivanç, Kiliç Cenk, Ozen Tuncer, Yüksel Selcen Pehlivan

机构信息

Faculty of Dentistry, Department of Oral Diagnosis and Radiology, Ankara University, Ankara, Turkey.

出版信息

Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2009 Feb;107(2):e34-42. doi: 10.1016/j.tripleo.2008.10.012.

Abstract

OBJECTIVE

The objective of this study was to assess the accuracy and reproducibility of cone-beam CT measurements of specific distances around the mandibular canal by comparing them to direct digital caliper measurements.

METHODS

Six formalin-fixed hemimandible specimens were examined using the ILUMA cone-beam CT system. Images were obtained at 120 kVp, 3.8 mA, and a voxel size of 0.2 mm, with an exposure time of 40 seconds. Specimens were cut into sections at 7 locations using a Lindemann burr, and a digital caliper was used to measure the following distances on both the anterior and posterior sides of each section: Mandibular Width (W); Mandibular Length (L); Upper Distance (UD); Lower Distance (LD); Buccal Distance (BD); and Lingual Distance (LID). The same distances were measured on the corresponding cross-sectional cone-beam CT images using the built-in measurement software. All caliper and cone-beam CT measurements were made by 2 independent trained observers and were repeated after an interval of 1 week. The Bland/Altman method was used to calculate intra- and inter-rater reliability. Intra-class correlation coefficients (ICCs) from 2-way random effects model were calculated. Agreements between cone-beam CT and direct digital caliper were calculated by ICC for 6 distances and 2 observers.

RESULTS

Intraobserver and interobserver measurements for all distances showed high agreement. ICCs for intraobserver agreement ranged from 0.86 to 0.97 for cone-beam CT measurements and from 0.98 to 0.99 for digital caliper measurements. ICCs between observers ranged from 0.84 to 0.97 for the cone-beam CT measurements and from 0.78 to 0.97 for the digital caliper measurements. ICCs for cone-beam CT and direct digital caliper ranged from 0.61 to 0.93 for the first observer and from 0.40 to 0.95 for the second observer.

CONCLUSION

Accuracy of cone-beam CT measurements of various distances surrounding the mandibular canal was comparable to that of digital caliper measurements.

摘要

目的

本研究的目的是通过将锥形束CT测量下颌管周围特定距离与直接数字卡尺测量结果进行比较,来评估其准确性和可重复性。

方法

使用ILUMA锥形束CT系统对6个福尔马林固定的半下颌骨标本进行检查。在120 kVp、3.8 mA的条件下获取图像,体素大小为0.2 mm,曝光时间为40秒。使用林德曼钻在7个位置将标本切成切片,并使用数字卡尺在每个切片的前后两侧测量以下距离:下颌宽度(W);下颌长度(L);上距离(UD);下距离(LD);颊侧距离(BD);和舌侧距离(LID)。使用内置测量软件在相应的锥形束CT横断面图像上测量相同的距离。所有卡尺和锥形束CT测量均由2名经过独立培训的观察者进行,并在间隔1周后重复测量。采用Bland/Altman方法计算观察者内和观察者间的可靠性。计算双向随机效应模型的组内相关系数(ICC)。通过ICC计算6个距离和2名观察者的锥形束CT与直接数字卡尺之间的一致性。

结果

所有距离的观察者内和观察者间测量均显示出高度一致性。锥形束CT测量的观察者内一致性ICC范围为0.86至0.97,数字卡尺测量的ICC范围为0.98至0.99。观察者之间的锥形束CT测量ICC范围为0.84至0.97,数字卡尺测量的ICC范围为0.78至0.97。第一名观察者的锥形束CT与直接数字卡尺的ICC范围为0.61至0.93,第二名观察者的ICC范围为0.40至0.95。

结论

锥形束CT测量下颌管周围各种距离的准确性与数字卡尺测量相当。