Progressive atrial lead perforation developed 5 years after pacemaker replacement.

机构信息

Department of Cardiology, St. Mary's Hospital, Kurume, Japan.

出版信息

J Cardiol. 2009 Feb;53(1):150-3. doi: 10.1016/j.jjcc.2008.07.006. Epub 2008 Aug 26.

Abstract

This report describes an asymptomatic case of atrial lead perforation which developed 5 years after pacemaker implantation. Although retrospective findings of computed tomography showed a screw-in atiral lead had already perforated 9 months after the implantation, the lead protrusion on chest X-rays and pacing failure had not been noticed until another 3 years later. At first, this complication was managed conservatively, however, a lead perforation progressively developed and, as a result, open surgery was performed to remove the lead. We therefore should be aware of the potential occurrence of a progressive protrusion of a perforated lead.

摘要

本报告描述了一例起搏器植入 5 年后发生的无症状性心房导线穿孔病例。尽管 CT 回顾性检查发现,在植入后 9 个月时,螺旋型心房导线已经穿孔,但直到 3 年后,才在胸部 X 线片上发现导线突出和起搏失败。起初,该并发症采用保守治疗,但随后导线穿孔逐渐进展,最终进行了开胸手术以取出导线。因此,我们应该意识到穿孔导线逐渐突出的潜在风险。