Arias Miguel A, Puchol Alberto, Pachón Marta, Castellanos Eduardo, Rodríguez-Padial Luis
Cardiac Arrhythmia and Electrophysiology Unit, Department of Cardiology, Hospital Virgen de la Salud, Planta Semisótano, Avda. Barber 30, 45004 Toledo, Spain.
Europace. 2009 Apr;11(4):535-6. doi: 10.1093/europace/eun394. Epub 2009 Jan 26.
We report on a 75-year-old woman who successfully underwent cardiac resynchronization therapy. Two weeks after implantation, right continuous pectoral stimulation appeared. Retraction and dislodgement of the right atrial and right ventricular leads as well as coiling of the three leads around the device generator were observed on chest radiography, but, surprisingly, the left ventricular lead remained unaffected due to its major length to facilitate the manoeuvers of implantation.