Suppr超能文献

Reel syndrome in an ICD with DF4 connector.

作者信息

Wollmann Christian G

机构信息

III. Med. Klinik, Universitätsklinikum St. Pölten-Lilienfeld, Propst Führer-Str. 4, 3100, St. Pölten, Austria,

出版信息

Herzschrittmacherther Elektrophysiol. 2014 Sep;25(3):206-8. doi: 10.1007/s00399-014-0323-9. Epub 2014 Jul 3.

Abstract

This is the first case report about a patient implanted with a DF4 lead suffering from reel syndrome. The reel syndrome may occur due to a ratchet mechanism where the device generator creates torque on a lead fixed at the fixation sleeve. Theoretically, this is less likely to occur in DF4 connectors due to their higher stiffness.