Tailor Ramesh, Ibbott Geoffrey, Lampe Stephanie, Warren Whitney Bivens, Tolani Naresh
Radiation Physics, UT MD Anderson Cancer Center, Houston, Texas 77030, USA.
Med Phys. 2008 Dec;35(12):5861-8. doi: 10.1118/1.3020754.
Dosimetry measurements of a 131Cs brachytherapy source have been performed in liquid water employing thermoluminescence dosimeters. A search of the literature reveals that this is the first time a complete set of dosimetric parameters for a brachytherapy "seed" source has been measured in liquid water. This method avoids the medium correction uncertainties introduced by the use of water-equivalent plastic phantoms. To assure confidence in the results, four different sources were employed for each parameter measured, and measurements were performed multiple times. The measured dosimetric parameters presented here are based on the AAPM Task Group 43 formalism. The dose-rate constant measured in liquid water was (1.063 +/- 0.023) cGy h(-1) U(-1) and was based on the air-kerma strength standard for this source established by the National Institute of Standards and Technology. Measured values for the 2D anisotropy function and the radial dose function are presented.