Helical microwave intracavitary oesophageal (HMIO) applicators were designed to operate at frequencies of 433 MHz and 915 MHz. Heating patterns were studied within muscle-equivalent phantom by thermographic camera and fibreoptic thermometers. The results showed that frequency significantly influenced the microwave heating pattern. The 433 MHz applicator had a single power deposition region, the longitudinal specific absorption rate (SAR) distribution appeared to be nearly even, and the maximum SAR value occurred close to the centre of the active length of the applicator. The 915 MHz applicator had two power deposition regions, the peak SAR values occurred at about 1/4 and 3/4 of the active length respectively. The radial SAR distribution suggested that there is no obvious difference between the 433 MHz and 915 MHz applicators in that the average radial penetration of 50% surface SAR (RP50) was about 0.65 cm. It was also shown that power deposition was axially symmetric for both 433 MHz and 915 MHz HMIO applicators. It is shown that better impedance matching is more important for intracavitary hyperthermia than for external hyperthermia. Choosing HMIO applicators in clinical practice is also discussed.
螺旋微波腔内食管(HMIO)施源器设计用于在433兆赫和915兆赫频率下运行。通过热成像相机和光纤温度计在肌肉等效体模内研究加热模式。结果表明,频率对微波加热模式有显著影响。433兆赫的施源器有一个单一的功率沉积区域,纵向比吸收率(SAR)分布似乎几乎均匀,最大SAR值出现在施源器有效长度中心附近。915兆赫的施源器有两个功率沉积区域,峰值SAR值分别出现在有效长度的约1/4和3/4处。径向SAR分布表明,433兆赫和915兆赫的施源器之间没有明显差异,因为50%表面SAR(RP50)的平均径向穿透约为0.65厘米。还表明,433兆赫和915兆赫的HMIO施源器的功率沉积都是轴对称的。结果表明,对于腔内热疗而言,更好的阻抗匹配比体外热疗更重要。文中还讨论了在临床实践中选择HMIO施源器的问题。