Rata Mihaela, Salomir Rares, Umathum Reiner, Jenne Jürgen, Lafon Cyril, Cotton François, Bock Michael
Inserm, U556, Université de Lyon, Lyon, F-69003, France.
Phys Med Biol. 2008 Nov 21;53(22):6549-67. doi: 10.1088/0031-9155/53/22/017.
High-intensity contact ultrasound (HICU) under MRI guidance may provide minimally invasive treatment of endocavitary digestive tumors in the esophagus, colon or rectum. In this study, a miniature receive-only coil was integrated into an endoscopic ultrasound applicator to offer high-resolution MRI guidance of thermotherapy. A cylindrical plastic support with an incorporated single element flat transducer (9.45 MHz, water cooling tip) was made and equipped with a rectangular RF loop coil surrounding the active element. The integrated coil provided significantly higher sensitivity than a four-element extracorporeal phased array coil, and the standard deviation of the MR thermometry (SDT) improved up to a factor of 7 at 10 mm depth in tissue. High-resolution morphological images (T1w-TFE and IR-T1w-TSE with a voxel size of 0.25 x 0.25 x 3 mm3) and accurate thermometry data (the PRFS method with a voxel size of 0.5 x 0.5 x 5 mm3, 2.2 s/image, 0.3 degree C voxel-wise SDT) were acquired in an ex vivo esophagus sample, on a clinical 1.5T scanner. The endoscopic device was actively operated under automatic temperature control, demonstrating a high level of accuracy (1.7% standard deviation, 1.1% error of mean value), which indicates that this technology may be suitable for HICU therapy of endoluminal cancer.
在MRI引导下的高强度接触式超声(HICU)可为食管、结肠或直肠的腔内消化肿瘤提供微创治疗。在本研究中,将一个微型接收线圈集成到内镜超声探头中,以提供热疗的高分辨率MRI引导。制作了一个带有内置单元素平面换能器(9.45 MHz,水冷尖端)的圆柱形塑料支架,并配备了围绕有源元件的矩形RF环形线圈。集成线圈的灵敏度明显高于四元素体外相控阵线圈,在组织10 mm深度处,MR温度测量的标准偏差(SDT)提高了7倍。在临床1.5T扫描仪上,在离体食管样本中获取了高分辨率形态图像(体素大小为0.25×0.25×3 mm3的T1w-TFE和IR-T1w-TSE)和准确的温度测量数据(体素大小为0.5×0.5×5 mm3、每幅图像2.2 s、体素级SDT为0.3℃的PRFS方法)。内镜设备在自动温度控制下积极运行,并显示出高度的准确性(标准偏差为1.7%,平均值误差为1.1%),这表明该技术可能适用于腔内癌症的HICU治疗。