Brett J L, Humble M W
Department of Laboratory Services, Wellington Hospital.
N Z Med J. 1991 Oct 9;104(921):422-4.
to compare the BACTEC TB system with our conventional multimedia system for the isolation of mycobacteria from clinical specimens.
specimens received over a two year period were cultured into BACTEC 12B vials and onto a variety of solid and liquid media routinely used by us for the isolation of mycobacteria. Cultures were observed regularly for bacterial growth, and examined for the presence of mycobacteria by acid fast microscopy. All results were recorded. At the end of the two year period the results for the two methods were compared.
of the 4443 specimens cultured by both methods, 260 were positive for mycobacteria. Sensitivity and specificity were 88.31% and 99.35% respectively for the BACTEC, and 84.41% and 99.96% respectively for our routine multimedia system. Tuberculosis was confirmed by the BACTEC TB system in six patients who were otherwise culture negative. One of these patients had tuberculous meningitis. Three patients culture positive for M tuberculosis by our routine method, were negative by the BACTEC. Twenty-nine false positive cultures were detected, 27 by the BACTEC TB system alone.
the BACTEC TB system is a rapid and sensitive system for the detection of mycobacteria in clinical specimens, and has proved to be a useful adjunct to our routine multimedia system. However, decreased specificity, particularly with lower respiratory tract specimens, means that positive results from these sites can be reported less confidently when they are detected by the BACTEC TB system alone. This negates some of the advantages of the BACTEC TB system for the detection of mycobacterial disease of the lungs.
比较BACTEC结核系统与我们的传统多媒体系统从临床标本中分离分枝杆菌的效果。
将两年期间收到的标本接种于BACTEC 12B瓶中,并接种到我们常规用于分离分枝杆菌的各种固体和液体培养基上。定期观察培养物的细菌生长情况,并通过抗酸显微镜检查分枝杆菌的存在。记录所有结果。在两年期结束时,比较两种方法的结果。
两种方法培养的4443份标本中,260份分枝杆菌呈阳性。BACTEC的敏感性和特异性分别为88.31%和99.35%,我们的常规多媒体系统的敏感性和特异性分别为84.41%和99.96%。BACTEC结核系统在另外6名培养结果为阴性的患者中确诊为结核病。其中1例患者患有结核性脑膜炎。我们常规方法培养的3例结核分枝杆菌阳性患者,BACTEC检测为阴性。检测到29例假阳性培养物,其中仅BACTEC结核系统检测到27例。
BACTEC结核系统是一种快速、敏感的检测临床标本中分枝杆菌的系统,已被证明是我们常规多媒体系统的有用辅助手段。然而,特异性降低,尤其是在下呼吸道标本中,这意味着仅通过BACTEC结核系统检测到这些部位的阳性结果时,报告的可信度较低。这抵消了BACTEC结核系统在检测肺部分枝杆菌病方面的一些优势。